Note: Any other height -50 away from the calculated optimum is mostly good. The
proper height number is most important when conducting half cell impedance measurements.
If you use a wrong plunger gap, you may get trouble with your battery test. The below figures illustrate
this point.
Good case: No bending of
built-in separator
Plunger height number too large:
Excessive downward bending of
built-in separator
Plunger height number too small:
Excessive upward bending of
built-in separator