PAT-Cell-Gas

 

Release 1.34

 

PAT-Cell-Gas SP 

 

 

 

4  Safety precautions 

Use proper safety precautions when using hazardous electrode materials and electrolytes. 
Wear protective glasses and gloves to protect you against electrolyte that may accidentally 
spill out during disassembly. Upon cell disassembly, dispose all materials properly. Metallic 
lithium and some insertion compounds may decompose heavily in contact with water and 
other solvents, and can cause fire.  

 

 

Features 

 

Gas inlet and outlet  

 

Gas sample port 

 

Gas pressure sensor, 0 to 3 bar abs 

 

Содержание PAT-Cell-Gas

Страница 1: ...2022 EL Cell GmbH User Manual Release 1 34 PAT Cell Gas Electrochemical test cell for in situ gas analysis in a flow through set up...

Страница 2: ...ies EL Cell GmbH maintains the right to make changes without further notice to products described in this manual to improve reliability function or design EL Cell GmbH does not assume any liability ar...

Страница 3: ...3 Variants 7 4 Safety precautions 8 5 Assembly of the PAT Cell Gas 9 6 Working modes of the PAT Cell Gas 11 7 Disassembly and cleaning 13 8 Sample valve with septum port options S and SP 14 9 Unpacki...

Страница 4: ...e a separator and the upper counter electrode The cell can be operated with or without reference electrode The gas permeable electrode is seated on a special lower plunger for current and gas feed Two...

Страница 5: ...Page 5 of 23 User Manual PAT Cell Gas Release 1 34 Schematic view Gas inlet and outlet Inlet Outlet...

Страница 6: ...1 Dead volume PAT Cell Gas with metal seal PAT Cell Gas without PAT Core 8 275 ml PAT Cell Gas with PAT Core Lower plunger AIR 4 638 ml PAT Cell Gas with PAT Core Lower plunger flow field 3 915 ml 2...

Страница 7: ...l PAT Cell Gas Release 1 34 3 Variants PAT Cell Gas PAT Cell Gas P PAT Cell Gas S Features Gas inlet and outlet Features Gas inlet and outlet Gas pressure sensor 0 to 3 bar abs Features Gas inlet and...

Страница 8: ...r protective glasses and gloves to protect you against electrolyte that may accidentally spill out during disassembly Upon cell disassembly dispose all materials properly Metallic lithium and some ins...

Страница 9: ...t cell A more detailed description of the PAT Core can be found in a separate manual Note The assembly has to take place under the protective atmosphere in a glove box Screw cap Sealing ring Upper plu...

Страница 10: ...de the cell base Then insert the assembly into the cell base 6 Evenly dispense approx 100 L of electrolyte on top of the separator with a pipette Note The optimum amount of electrolyte will depend on...

Страница 11: ...rrent collector supporting it During operation the pressure gradient building up between cell headroom and the gas volume below the perforated plate effectively prevents back mixing The relatively lar...

Страница 12: ...eld below the working electrode The composition of the outgoing gas can be analyzed by e g mass spectrometry The pressure gradient between cell headroom and spiral type flow field effectively prevents...

Страница 13: ...n particular avoid contact of the electronic components on the bottom of the cell base with liquid For the cell variants with pressure sensor options P and SP the two adjacent holes at the top of the...

Страница 14: ...e subsequent gas analysis with e g a gas chromatograph We recommend using a syringe with a pencil point needle in order to prevent clogging when the septum of the sample port is pierced A 1 ml syringe...

Страница 15: ...be purchased separately Sealing ring PEEK 5x ECC1 00 0232 D V Sealing ring PE 10x ECC1 00 0232 A X PAT Cell Gas P PAT Cell Gas test cell without PAT Core Note The components of the PAT Core insulation...

Страница 16: ...re Note The components of the PAT Core insulation sleeves and plungers must be purchased separately Septum 10x ECC1 00 0097 B X Syringe 1ml w o Luerlock LAB0024 Spinal needle pencil point 27g LAB0039...

Страница 17: ...are parts and consumables PAT Cell Gas SP complete Screw cap insulated PAT complete ECC1 00 0236 D Valve block I PAT Gas assy ECC1 00 0555 A O Ring 3mm x 1 5mm AP370 DIC9059 Cell bases divided GAS ass...

Страница 18: ...Page 18 of 23 User Manual PAT Cell Gas Release 1 34 Screw cap insulated PAT complete ECC1 00 0236 D...

Страница 19: ...Page 19 of 23 User Manual PAT Cell Gas Release 1 34 Cell base divided GAS assy ECC1 00 0530 E Cell bases divided GAS assy ECC1 00 0530 E O Ring 3mm x 1 5mm AP370 DIC9059...

Страница 20: ...k I PAT Gas assy ECC1 00 0555 A O Ring 3mm x 1 5mm DIC9059 Adapter flange 1 16 male welded ECC1 01 0071 C O Ring 1 5mm x 2mm DIC9007 Valve body IV ECC1 00 0391 D Nut for 1 16 in Swagelok Tube Fitting...

Страница 21: ...Page 21 of 23 User Manual PAT Cell Gas Release 1 34 Gas sample port PAT assy ECC1 00 0155 C...

Страница 22: ...at the Seller s option demand return of the goods and tender repayment of the price Buyer s exclusive remedy is repair or replacement of defective and nonconforming goods or at Seller s option the re...

Страница 23: ...f all other warranties expressed or implied including the warranties of merchantability and fitness for a particular purpose This Limited Warranty supersedes all prior proposals or representations ora...

Отзывы: