5 Assembly and Connection
The test cell can either be operated in 2-electrode or 3-electrode (reference)
configuration. In the following, the cell assembly is described for operation with a
lithium metal reference electrode. For more details, and for 2-electrode operation,
please refer to the separate manuals of the ECC-Ref and the ECC-Std test cell.
Generally, all assembly steps are to be carried out in inert glove box atmosphere,
and all components used are to be dried upfront in a vacuum oven at 80°C for at
least 12 hours. The pressure sensor together with the transfer line may remain
permanently attached to the stand.
Assembly:
i)
Load the feed-through hole of the Ref sleeve with a small amount of lithium
metal with the aid of a metal wire. A dedicated tool is optionally available to
ease the procedure (
https://el-cell.com/products/tools-accessories/ecc-
refload
).
ii)
Insert the Ref sleeve into the cell base.
iii)
Mount the locking ring, thereby aligning the side opening of the cell base with
the feed-through hole of the Ref sleeve.
iv)
Attach the reference pin assembly to the cell base.
v)
Place the counter electrode (e.g. a lithium metal disc, 0.2 mm thick, 18 mm
diameter) into the sleeve
vi)
Put the glass fibre separator on top.
vii)
Dispense a defined amount of electrolyte onto the separator, and place the
working electrode, with its active layer downside, on top inside the sleeve.
viii)
Insert the plunger, the lid seal and the spring.
NOTE
: Two types of lid seals
are provided made of LDPE and PEEK, respectively. The PEEK seals are
recommended for use at elevated temperatures and for lowest leakage rates.
Lid seals are single-use components. Re-use may result in increased leakage.
ix)
Optionally, insert the dead volume spacer into the spring.
NOTE
: The smaller
the dead volume, the higher the pressure change
p/
V for a given amount of
gas.
x)
Attach the cell lid, insert the cell into the bracket, and tighten the wing nut.
xi)
Connect the cell towards the pressure sensor by attaching the transfer line to
the sensor port of the cell.