Eko Devices DUO E5 Скачать руководство пользователя страница 2

P.

2

The Eko Model E5 System is designed to support 
healthcare professionals in analyzing cardiac and 
other internal organ sounds. The Eko Model E5 
System includes a device capturing heart sounds 
and ECG readings (herein referred to as DUO), a 
smartphone application (App), and a web application 
(Dashboard).

DUO features audio and ECG data transmission via 
Bluetooth that allows the user to open and playback 
sounds and visualize phonocardiograms (PCG) and 
electrocardiograms (ECG) in a mobile application on 
compatible iOS smartphones and tablets. The App 
provides the ability to save audio recordings and 
waveforms (PCG and ECG) within Electronic Health 
Record (EHR) systems, share patient recordings with 
other practitioners, and annotate notes on recorded 
audio. Eko is intended for use on pediatric and adult 
patients. The device is not intended for infants 
weighing less than 10 kg.

CAUTION:

 Federal (USA) law restricts this device to 

sale to or on the order of a clinician.

For general and product related comments, questions, 
or concerns, please contact Eko Devices, Inc. directly

Please report any injury or adverse event to Eko using 
any of the contact methods below.

Eko Devices, Inc.

2600 10th St. Suite 260
Berkeley, CA 94710
USA

General Assistance and FAQs

ekodevices.com/getstarted

Direct Contact

[email protected]

Phone Support

1.844.356.3384

Product Reference and Information

www.ekodevices.com

1.1

Introduction, Warnings, and Safety

1.2

For Help and Assistance

Consult instructions for use.

This product contains electrical and 
electronic components and must not be 
disposed of using standard refuse collection. 
Please consult local directives for disposal of 
electrical and electronic equipment.

This product and packaging does not contain 
natural rubber latex.

This product contains an intentional RF 
radiator certified by the FCC.

Catalog Number

Lot Number

Serial Number

Humidity Limit (Operational)

Temperature Limit (Operational)

This product is provided non-sterile.  Do not 
attempt to re-sterilize the device.

This product uses wireless Bluetooth 
communication

BF Applied Part

Contents (Quantity)

Protection against dust and water jets when 
earpieces are attached. 
The earpieces are not IP protected.

MR Unsafe

1.3

Safety Related Labels & Symbols

LATEX

10%

95%

45° C

113° F

10° C

50° F

NON

STERILE

SN

LOT

REF

IP55

Содержание DUO E5

Страница 1: ...User Manual Model E5...

Страница 2: ...es Inc directly Please report any injury or adverse event to Eko using any of the contact methods below Eko Devices Inc 2600 10th St Suite 260 Berkeley CA 94710 USA General Assistance and FAQs ekodevi...

Страница 3: ...y move DUO away from that device and or turn the Bluetooth feature OFF To reduce the risks associated with environmental contamination follow applicable regulations when disposing of this device DUO c...

Страница 4: ...rence and 2 this device must accept any interference received including interference that may cause undesirable operation NO MODIFICATION Modifications to this device shall not be made without the wri...

Страница 5: ...but not limited to patient body temperature and clean contact area The device uses a Bluetooth Class 2 wireless data link The Bluetooth range will be reduced when objects walls furniture people etc a...

Страница 6: ...isclosure of any electronic patient data through the use of software If the user becomes unable to comply with a law or restriction that applies to use and disclosure of such data the user should not...

Страница 7: ...power adapter The LED light ring on DUO will begin to flash to signal the charging status All of the LEDs in the ring will light up when the device is fully charged DUO should be periodically recharge...

Страница 8: ...ted on the top face An LED ring illuminates when the device is powered on Power Save Mode DUO will automatically power off after 5 minutes of continuous inactivity To resume DUO press and hold the act...

Страница 9: ...ing light up based on volume level The device is ON The filter change is indicated by the active LED in the LED ring rotating to a new position The device is ON and has low battery The red warning LED...

Страница 10: ...dio press the device firmly against the skin to ensure good contact ECG Capture ECG by placing and orienting DUO on the chest The optimal position to capture ECG data is at an inward facing 45 angle a...

Страница 11: ...e device and may affect battery life Operating Warnings Failure to follow care and maintenance recommendations could result in damage to the internal components of the device Internal damage to the pr...

Страница 12: ...4 digit login PIN to simplify the login process on the Eko mobile app To create or edit a login PIN manually select the Menu icon on the top left of the home screen Select Account Settings PIN and fol...

Страница 13: ...top right of the home screen Select the desired patient and select a recording to add notes to Scroll to the bottom of the recording screen Select Add Comment and begin typing your note Select the th...

Страница 14: ...otes 25mm s and 10mm mV 7 3 Main Recording Screen 89 AT T 1 45 PM Hill Thomas Settings Modify Bluetooth Adjust Volume Change Audio Source iPhone speakers or DUO Earpieces Adjust Account Details Name U...

Страница 15: ...PIN is recommended so that users do not have to enter full login information each time the App is opened Finally users may log out of the App from this screen 7 4 Settings Screen My Devices Audio Sett...

Страница 16: ...ted with the patient Patients may be added to the list individually through manual input or syncing with compatible EHR services 7 5 Patients Screen Aaberg Mark Anonymous PS01CEP90784 Aortic stenosis...

Страница 17: ...ok at 01 57 for me Martin Stemper 4 59pm RRR Murmur heard and Indicated Martin Stemper 4 59pm This is one of the more complicated afib cardiology characteristics Will address tomorrow You 4 59pm NOTES...

Страница 18: ...ation in the configuration in which it will be used Guidance and Manufacturer s Declaration Electromagnetic Emission DUO is intended for use in the electromagnetic environment specified below The user...

Страница 19: ...z 3 Vrms 150kHz 80 Mhz Portable and mobile RF communications equipment should be used no closer to any part of DUO including cables than the recommended separation distance calculated from the equatio...

Страница 20: ...rding to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 2 1 2 2 3 10 3 7 3 7 7 4 100 12 12 23 For...

Страница 21: ...www ekodevices com 2016 Eko Devices Inc All Rights Reserved Eko the Eko logo and DUO are trademarks of Eko Other company and product names may be trademarks of their respective owners LBL 009 B...

Отзывы: