Electromagnetic Compatibility
Like other electrical medical equipment, the
EX190W
requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To ensure
electromagnetic compatibility (EMC), the EX190W must be installed and operated according
to the EMC information provided in this manual.
Note
The EX190W has been designed and tested to comply with IEC 60601-1-
2:2001 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the EX190W����
Warning
Do not use cables or accessories other than those provided
with the EX190W, as this may result in increased electromagnetic emissions
or decreased immunity to such emissions��
Warning If the EX190W is used adjacent to or stacked with other
equipment, observe and verify normal operation of the EX190W in the con-
figuration in which it will be used prior to using it in a surgical procedure��
Consult the tables below for guidance in placing the EX190W��
The Model EX190W is intended for use in the electromagnetic environment specified below.
The customer or the user of EX190W should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The Model EX190W uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment
RF emissions CISPR 11
Class B
EX190W is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes
Harmonic emissions
IEC61000-3-2
Class D
Voltage Fluctuations
IEC61000-3-3
Complies
Manufacturer’s declaration - electromagnetic emission
The Model EX190W is intended for use in the electromagnetic environment specified below.
The customer or the user of EX190W should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient
/burst IEC61000-4-4
2 kV for power supply lines
1 kV for input / output lines
2 kV for power supply lines
1 kV for input / output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Manufacturer’s declaration - electromagnetic immunity
EX190W
29
ENGLISH