Recommended separation distances between portable or mobile RF communication equipment
and the EX4342
The EX4342 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the EX4342 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the EX4342.
Immunity to proximity fields from the following RF wireless communication equipment has been confirmed:
Test
frequency
(MHz)
Bandwidth
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Minimum
separation
distance
(m)
IEC /
EN60601
test level
(V/m)
Compliance
level
(V/m)
385
380 - 390
TETRA 400
Pulse modulation
b)
18 Hz
1.8
0.3
27
27
450
430 - 470
GMRS 460,
FRS 460
FM
±5 kHz deviation
1 kHz sine
2
0.3
28
28
710
704 - 787
LTE Band 13, 17 Pulse modulation
b)
217 Hz
0.2
0.3
9
9
745
780
810
800 - 960
GSM 800 / 900,
TETRA 800,
iDEN 820
CDMA 850,
LTE Band 5
Pulse modulation
b)
18 Hz
2
0.3
28
28
870
930
1720
1700 - 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3, 4,
25;
UMTS
Pulse modulation
b)
217 Hz
2
0.3
28
28
1845
1970
2450
2400 - 2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
b)
217 Hz
2
0.3
28
28
5240
5100 - 5800
WLAN 802.11 a/n Pulse modulation
b)
217 Hz
0.2
0.3
9
9
5500
5785
a) For some services, only the uplink frequencies are included.
b)
Carrier waves are modulated using a 50 % duty cycle square wave signal.
The EX4342 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
For other portable and mobile RF communication equipment (transmitters), the minimum distance between portable
and mobile RF communications equipment (transmitters) and the EX4342 should be according to the maximum output
power of the communication equipment as recommended below.
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.7 GHz
d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance “d” in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where “P” is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1
At 80 MHz and 800 MHz, the separation distance for a higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflec
-