125
3.
If a warning that the air detection is disabled (OFF) appears, ensure that a set with
an air-eliminating filter is used, and press
OK
.
4.
If the
Dosing method
screen appears, select the appropriate option:
•
mL:
.
•
Dose calculation:
.
Weight based units are available for both Dosing methods.
Otherwise, continue to Step
5.
From the
Concentration units
screen, select the appropriate drug units.
6.
According to pump configuration one of the following screens will appear:
•
Concentration
: From the
Concentration
screen, using the keypad, enter the
Concentration
OK
•
Drug amount
: Using the keypad, enter the
Drug Amount
OK
. Then, using
the keypad, enter the
Diluent Volume
OK
7.
From the
Attention
screen, confirm the concentration and press
OK
.
If the selected drug unit is
Million Units
.
8.
If the
Patient Weight
screen appears, specify whether the infusion is weight based:
•
Yes
.
•
No
9.
From the
Patient Weight
screen, using the keypad, enter the
Patient Weight
OK
.
10.
From the
Dose Rate units
screen, select the appropriate dose rate units.
11.
Using the keypad, enter the number of Steps required for the infusion, and then
press
OK
.
12.
For the first Step, program 2 of the following 3 parameters, by selecting the
relevant boxes (the digit refers to the number of the Step):
•
Rate 1:
Using the keypad, enter the value
OK
.
To display additional concentration units press
Next
.
Содержание Q Core Sapphire
Страница 1: ......
Страница 68: ...Sapphire Infusion Pump User Manual 67 This page is left intentionally blank ...
Страница 82: ...Sapphire Infusion Pump User Manual 81 This page is left intentionally blank ...
Страница 188: ...Sapphire Infusion Pump User Manual 187 This page is left intentionally blank ...
Страница 198: ...Sapphire Infusion Pump User Manual 197 This page is left intentionally blank ...
Страница 244: ...Sapphire Infusion Pump User Manual 243 This page is left intentionally blank ...
Страница 270: ...Sapphire Infusion Pump User Manual 269 This page is left intentionally blank ...
Страница 288: ...Sapphire Infusion Pump User Manual 287 ...
Страница 292: ...Sapphire Infusion Pump User Manual 291 This page is left intentionally blank ...
Страница 293: ......
Страница 294: ......