Electrical Adjustment
Grp / No. Item
Function
Range
Initial
Note
5
Clog Auto Coolin Enable
Auto Cleaning at Clogged (0: Disable 1: Eanble)
0 - 1
1
6
Scroll Fail Auto Cooling Enable
Auto Cleaning at Scroll Error (0: Disable 1: Eanble)
0 - 1
1
7
Timer Force Scroll Enable
Forced Scroll when Filter Timer reached
0 - 1
1
241
Flow Ideal Data
(Value will be set after calibration)
0
Flow Ideal Data 1
Flow Sensor Theory Value No.1 Corresponding to Fan Volt No.1
0 - 1023
301
1
Flow Ideal Data 2
Flow Sensor Theory Value No.2 Corresponding to Fan Volt No.2
0 - 1023
343
2
Flow Ideal Data 3
Flow Sensor Theory Value No.3 Corresponding to Fan Volt No.3
0 - 1023
380
3
Flow Ideal Data 4
Flow Sensor Theory Value No.4 Corresponding to Fan Volt No.4
0 - 1023
400
4
Flow Ideal Data 5
Flow Sensor Theory Value No.5 Corresponding to Fan Volt No.5
0 - 1023
410
5
Flow Ideal Data 6
Flow Sensor Theory Value No.6 Corresponding to Fan Volt No.6
0 - 1023
1023
6
Flow Ideal Data 7
Flow Sensor Theory Value No.7 Corresponding to Fan Volt No.7
0 - 1023
1023
7
Flow Ideal Data 8
Flow Sensor Theory Value No.8 Corresponding to Fan Volt No.8
0 - 1023
1023
242
Flow Calib. Fan Volt
(Not Memorized)
0
Flow Calib Fan Volt 1
Flow Sensor Calibration Fan Voltage No.1 [x10 V]
0 - 255
50
1
Flow Calib Fan Volt 2
Flow Sensor Calibration Fan Voltage No.2 [x10 V]
0 - 255
70
2
Flow Calib Fan Volt 3
Flow Sensor Calibration Fan Voltage No.3 [x10 V]
0 - 255
95
3
Flow Calib Fan Volt 4
Flow Sensor Calibration Fan Voltage No.4 [x10 V]
0 - 255
120
4
Flow Calib Fan Volt 5
Flow Sensor Calibration Fan Voltage No.5 [x10 V]
0 - 255
138
5
Flow Calib Fan Volt 6
Flow Sensor Calibration Fan Voltage No.6 [x10 V]
0 - 255
255
6
Flow Calib Fan Volt 7
Flow Sensor Calibration Fan Voltage No.7 [x10 V]
0 - 255
255
7
Flow Calib Fan Volt 8
Flow Sensor Calibration Fan Voltage No.8 [x10 V]
0 - 255
255
243
Flow Calib Effective Difference
(Not Memorized)
0
Flow Calib Effective Diff 1
Flow Sensor Calibration Range Diff No.1 [ADC]
0 - 1023
180
1
Flow Calib Effective Diff 2
Flow Sensor Calibration Range Diff No.2 [ADC]
0 - 1023
180
2
Flow Calib Effective Diff 3
Flow Sensor Calibration Range Diff No.3 [ADC]
0 - 1023
180
3
Flow Calib Effective Diff 4
Flow Sensor Calibration Range Diff No.4 [ADC]
0 - 1023
180
4
Flow Calib Effective Diff 5
Flow Sensor Calibration Range Diff No.5 [ADC]
0 - 1023
180
5
Flow Calib Effective Diff 6
Flow Sensor Calibration Range Diff No.6 [ADC]
0 - 1023
1023
6
Flow Calib Effective Diff 7
Flow Sensor Calibration Range Diff No.7 [ADC]
0 - 1023
1023
7
Flow Calib Effective Diff 8
Flow Sensor Calibration Range Diff No.8 [ADC]
0 - 1023
1023
244
Flow Clog Threshold
(Not Memorized )
0
Flow Clog Threshold 1
Flow Sensor Clogged Threshold No.1[ADC]
0 - 1023
40
1
Flow Clog Threshold 2
Flow Sensor Clogged Threshold No.2[ADC]
0 - 1023
46
2
Flow Clog Threshold 3
Flow Sensor Clogged Threshold No.3[ADC]
0 - 1023
56
3
Flow Clog Threshold 4
Flow Sensor Clogged Threshold No.4[ADC]
0 - 1023
77
4
Flow Clog Threshold 5
Flow Sensor Clogged Threshold No.5[ADC]
0 - 1023
84
5
Flow Clog Threshold 6
Flow Sensor Clogged Threshold No.6[ADC]
0 - 1023
1023
6
Flow Clog Threshold 7
Flow Sensor Clogged Threshold No.7[ADC]
0 - 1023
1023
7
Flow Clog Threshold 8
Flow Sensor Clogged Threshold No.8[ADC]
0 - 1023
1023
245
Filter Roll
0
Take Up Bend
Remove Filter Flexure Starts when the Value is set to 1.
0 - 1
0
1
Take Up Times
Remaining Times of Filter Scroll (Read only)
-
-
2
Take Up Time Reset
Filter Scroll Times Reset
Resets Remaining Scroll Times when the Value is set to 10.
0 - 10
0
3
Scroll Counter Reset Time
Scroll Counter Reset Times (Read only)
0 - 255
-
4
Scroll Information Cartridge No
0 - 2
0
246
Flow Clog Monitor
(Read only)
0
Real Flow Data
Real Flow Data [ADC] (Read only)
0 - 1023
-
1
Ideal Flow Data
Theory Flow Data (Read only)
0 - 1023
-
2
Flow Difference Data
Flow Difference Data [ADC] (Read only)
0 - 1023
-
3
Clog Detect Flow Difference Data Difference Data to judge the Filter Clogged [ADC] (Read only)
0 - 1023
-
4
Warn Detect Flow Difference Data Difference Data to judge the Filter Clogged Warning [ADC] (Read
only)
0 - 1023
-
5
Press Add Data
Add Data by Air Pressure (Read only)
0 - 1023
-
6
Filter Scroll Timer
Timer for Filter Scroll [hour] (Read only)
0 - 32767
-
247
Filter Scroll Time
(Read only)
* 9999 means not executed
Содержание LC-HDT700
Страница 94: ...94 AD8183 Amp SW IC2001 IC5001 BA6287 Motor Drive Focus Zoom Shutter IC5501 IC5521 IC5561 IC Block Diagrams...
Страница 95: ...95 IC Block Diagrams BA6920 Motor Drive Lens Shift Up Down Left Right IC6501 IC6521 BA7078 Sync Separator 1011...
Страница 97: ...97 IC Block Diagrams ISL51002 A D Converter IC2201 M62393 DAC IC7801...
Страница 98: ...98 IC Block Diagrams MXA2500 Dual Axis Accelerometer IC1881 IC1882 NJM2671 Motor Controller IC5541...
Страница 99: ...99 IC Block Diagrams PW392 Scaler IC301 PW610 Keystone IC2301...
Страница 100: ...100 IC Block Diagrams TE7783 I O Expander IC4801...
Страница 103: ...103 Mechanical Parts List KY2 HDT70000 Shutter assembly M08...
Страница 149: ...Key No Part No Description Key No Part No Description KY2 HDT70000 149 Electrical Parts List...
Страница 150: ...KY7BL Nov 2011 Japan 2011 Eiki Internatinal Inc A key to better communications...
Страница 164: ...A12 SCH_KY2AL 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 A B C D E F G H I J K L A B C D E F G H I J K L...