Aquameter
™
Series Instruction Manual
10401-00188 Rev M
23.5. Operation
Connect the Flowcell to a pumping device so that sample water enters at the bottom and
exits at the top. Adjust the flow rate so that there is no visible turbulence or cavitation within
the Flowcell. Connect an Aquameter and monitor the readings. If the readings are jumpy or
erratic, reduce the flow rate. The ideal flow rate is around 30 litres/hour (8 US gallons/hour),
although the Aquaprobe is capable of operating at flow rates as low as 15 litres/hour (4 US
gallons/hour). Flow rates above 60 litres/hour (16 US gallons/hour) are not recommended.
23.6. Caution
The maximum operating pressure of the Flowcell is 300mb (4.4 PSI). Select your pumping
device accordingly. If necessary, use a three-way bypass valve so that this limit is not
exceeded.
23.7. Cleaning
After use, rinse the Flowcell thoroughly with fresh water. To remove stubborn deposits,
scrub the inside of the Flowcell with a bottlebrush and non-abrasive detergent, then rinse
thoroughly.
Never clean the Flowcell with concentrated acid or alkaline based cleaning products
such as Decon 90. These products can strip the anodised finish from the Flowcell
and damage the plastic components.
23.8. Flowcell Troubleshooting
Problem
Cause / Remedy
DO readings are
abnormally high or are
fluctuating wildly.
Aeration of sample water. Check all joints for air leaks.
Reduce flow rate to avoid cavitation.
Turbidity readings are
abnormally high.
Air bubbles adhering to the Turbidity Electrode lenses.
Agitate Flowcell to dislodge.
Aeration of sample water. Check all joints for air leaks.
Reduce flow rate to avoid cavitation.
Sample water is leaking
from around the top of the
screw collar.
Screw collar is not tight enough. Tighten up.
Grease / PTFE tape has not been applied to the joint
between the Probe Sleeve and Probe Body (single part
sleeves only). See ‘Aquaprobe Preparation Prior to First
Use’ on previous page.
Operating pressure is too high. Reduce pressure / flow rate.
Probe is forced up out of
the Flowcell during use.
Operating pressure is much too high. Reduce pressure /
flow rate.
© 2011 Aquaread Ltd.
www.aquaread.co.uk
Page 63 of 64