Parameters
bbe AlgaeTorch Manual
Page 101
Parameters
Access
Level
Name
Dimensio
n
Meaning
User
Serial number
Serial number of the hardware
User
Software version (major)
User
Software version (minor)
User
LED warmup time
s
Time to warm up the LEDs before
a measurement takes place
User
Measurement time
s
Duration of one measurement
User
Measurement interval
s
User
Cellcountmode
Switches between “µg Chl A/l” and
“cells/l”
Service
Wavelength LED0
nm
For information only
Service
Wavelength LED1
nm
For information only
Service
Wavelength LED2
nm
For information only
Service
Wavelength LED3
nm
For information only
Service
Measuring type
Switches between single,
continuous and interval
measurement
Service
Fingerprint LED 0 Cyanos
digits/µg/l
Result of the calibration of cyanos
Service
Fingerprint LED 1 Cyanos
digits/µg/l
Result of the calibration of cyanos
Service
Fingerprint LED 2 Cyanos
digits/µg/l
Result of the calibration of cyanos
Service
Fingerprint LED 3 Cyanos
digits/µg/l
Result of the calibration of cyanos
Service
Fingerprint LED 0 green
algae
digits/µg/l
Result of the calibration of green
algae
Service
Fingerprint LED 1 green
algae
digits/µg/l
Result of the calibration of green
algae
Service
Fingerprint LED 2 green
algae
digits/µg/l
Result of the calibration of green
algae
Service
Fingerprint LED 3green
algae
digits/µg/l
Result of the calibration of green
algae
Service
Offset LED 0
digits
Signal of pure water
Service
Offset LED 1
digits
Signal of pure water
Service
Offset LED 2
digits
Signal of pure water
Service
Offset LED 3
digits
Signal of pure water
Service
Cellcount factor Cyanos
cells/µg/l
Factor to convert µg/l in cells /l
Service
Cellcount factor green
algae
cells/µg/l
Factor to convert µg/l in cells /l
Service
Global correction factor
1/100
Factor to adapt the results to a
local reference method
Service
Gradient turbidity
1/10
FTU/digit
Result of turbidity calibration
Service
Offset pressure
1/100 digits
Result of pressure calibration
Service
Gradient pressure
1/1000000
mbar/digits
Result of pressure calibration
Expert
req. val. LED 0 [nm]
digits
For brightness control
Expert
req. val. LED 1 [nm]
digits
For brightness control
Expert
req. val. LED 2 [nm]
digits
For brightness control
Expert
req. val. LED 3 [nm]
digits
For brightness control
Expert
Name algae class 0
Expert
Name algae class 1
Содержание 18.65.01
Страница 7: ...What s new bbe AlgaeTorch Manual Page 6 Dimensions 113...
Страница 43: ...bbe Software bbe AlgaeTorch Manual Page 42 Menu structure Basic functionality...
Страница 44: ...bbe Software bbe AlgaeTorch Manual Page 43 Menu structure Advanced functionality...
Страница 98: ...The Menus bbe AlgaeTorch Manual Page 97 X only applied Y only applied...
Страница 106: ...The GPS module bbe AlgaeTorch Manual Page 105...
Страница 114: ...Data and Tables bbe AlgaeTorch Manual Page 113 Dimensions...
Страница 115: ...Data and Tables bbe AlgaeTorch Manual Page 114...