11
Technical Data
Page 22 / 30
11.Technical Data
Manufacturer
Changzhou Sifary Medical Technology
Co., Ltd
Model
MotorTurbo
Power adapter input
AC 100~240V,50/60Hz 1.5A
Power adapter output
27V
2.5A
Cooling air flow
≤40L/min(0.25MPa-0.5MPa)
Spray air flow
≥1.5L/min(0.25MPa-0.5MPa)
Spray water flow
≥50mL/min(0.2MPa-0.5MPa)
Frequency
50/60Hz,±1Hz
Speed range
2000-40000RPM,±10%
Torque
>1N·cm
Handpiece connector
type
Comply with YY 1012:2004 handpiece
requirements
Light
White LED
Type of protection
against electrical shock
Class Ⅱ and internally powered
equipment
Applied part
B
Operation mode
Intermittent operation, working for 5
minutes / stopping for 2 minutes
Ingress Protection
IPX0
Operation conditions
Use: in enclosed spaces
Ambient temperature: 5°C ~ 40 °C
Relative humidity: <80%
Operating altitude < 3000m above sea
level
Transport and storage
conditions
Ambient temperature: -20 °C ~ +55 °C
Relative humidity: 20% ~ 80 %
Atmospheric pressure: 70kPa ~ 106kPa