Dental low-voltage Electrical Motor

USER MANUAL

Changzhou Sifary Medical Technology Co., Ltd.

MotorTurbo

Содержание MotorTurbo

Страница 1: ...Dental low voltage Electrical Motor USER MANUAL Changzhou Sifary Medical Technology Co Ltd MotorTurbo...

Страница 2: ...P N IFU 6435004 Version S01 Issued Oct 15 2021 Size 87mm X 180mm...

Страница 3: ...d removal of motor 8 4 3 Installation and removal of handpiece 9 5 Use Interface 10 5 1 Host operation panel 10 5 2 Screen display 10 5 3 Terms and definition 11 6 Setting 12 6 1 Set mode parameter 12...

Страница 4: ...on front plug Motor disinfection back plug 3 4 6 5 Note This product does not contain Contra angle handpiece 1 2 Components and Accessories Host 1pcs Motor 1pcs Power cable 1pcs Power Adapter 1pcs Mot...

Страница 5: ...rial number Catalogue number Manufacturer Date of manufacture LOT Batch number Safety class II device Type B applied part Direct current Dispose of in accordance with the WEEE directive Keep dry Steri...

Страница 6: ...equipment in the presence of free oxygen anesthetic gas or combustible materials The equipment must be operated used and stored in a safe environment 4 The device requires special precautions with reg...

Страница 7: ...he operation of the main engine otherwise the contra angle and motor will be damaged 10 Turn off the power switch After each use 11 Heavy hit such as falling will cause damage to the dental low voltag...

Страница 8: ...en the nut Connect the power adapter to the power connector and connect the power cable to power adapter then connect the power cable plug into a power source that meets the requirements refer to 11 T...

Страница 9: ...allel from the motor WARNING 1 Before install the handpiece make sure that the O ring on the motor connector is lubricated If there is no lubricant please spray lubricant special lubricant for the han...

Страница 10: ...each working mode Rotation direction button short press to switch the motor forward reverse the host will have a continuous and slow di di sound when reversing Rotation speed increase button increase...

Страница 11: ...eversal indicates the motor is reversing in the current working mode Dental handpiece output speed indicates the current output speed of the dental handpiece The range of rotation speeds that can be s...

Страница 12: ...lighted Set rotation 200000RPM rotation range 10000RPM 200000RPM step value is 5000RPM M2 mode parameter setting connected to a handpiece with a transmission ratio of 1 1 Set the working mode M2 Set t...

Страница 13: ...transmission ratio of 20 1 Set the working mode M4 Set the transmission ratio 20 1 Set rotation direction FWD Set LED light status highlighted Set rotation 100RPM rotation range 100RPM 2000RPM step va...

Страница 14: ...of dental treatment machine calibration has been carried out 2 Before using it for treatment please try it outside the oral cavity to make sure that there is no problem with the function of the device...

Страница 15: ...n In case of damage the device should be reprocessed before sending back to the manufacturer for repair 8 2 General recommendations The user is responsible for the sterility of the product for the fir...

Страница 16: ...gents glutaral etc or any other special types of water or commercial cleaning liquids Such liquids may result in metal corrosion and adhesion of the residual medical agents to the components Do not ri...

Страница 17: ...n To prevent the steam from entering the motor please screw the disinfection front and back plug to motor two side before sterilization Refer to the figure below Completed state WARNING 1 Be sure to s...

Страница 18: ...6 Wait for cooling before touching Storage Storage of sterilized instruments in a dry clean and dust free environment at modest temperatures refer to label and instructions for use WARNING 1 Sterilit...

Страница 19: ...h a cloth lightly moistened with Ethanol for Disinfection Ethanol 70 to 80vol at least 2min repeat for 5 times NOTE 1 Do not use disinfectants other than alcohol for disinfection 2 Do not use excessiv...

Страница 20: ...is time though operate the control foot switch the host will not work Adjust the input pressure to the range of 0 25 0 5MPa If the fault light is still on please contact your local dealer When the mot...

Страница 21: ...s not plugged into the socket Please check the connection The socket is not energized Please check the connection The motor does not rotate The compressed air pressure is too low or the air resource i...

Страница 22: ...Torque 1N cm Handpiece connector type Comply with YY 1012 2004 handpiece requirements Light White LED Type of protection against electrical shock Class and internally powered equipment Applied part B...

Страница 23: ...ence in nearby electronic equipment RF emissions CISPR 11 Class B The MotorTurbo is suitable for use in all establishments incl uding domestic establishents and those directly connect ed to the public...

Страница 24: ...t Surge IEC 61000 4 5 Line to line 0 5kV 1kV Line to line 0 5kV 1kV Mains power quality should be that of a typical commercial or hospital environment Voltage dips IEC 61000 4 11 Voltage interruptions...

Страница 25: ...nvironment specified below The customer or the user of the MotorTurbo should assure that it is used in such an environment Immunity test IEC 60601 test level Complianc e level Electromagneti c environ...

Страница 26: ...used When they are used in close proximity to medical equipment and or systems the medical equipment and or systems basic safety and essential performance may be affected The MotorTurbo has been teste...

Страница 27: ...DMA 850 LTE Band 5 Pulse modu l ation 18Hz 2 0 3 28 870 930 1720 1700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modu l ation 217Hz 2 0 3 28 1845 1970 2450 2400 2 570 Blueto o...

Страница 28: ...f MotorTurbo and result in improper operation Cable information Cable Name Cable Length m Shielded or not Remark Adapter output Cable 1 No Power cable 1 5 No Motor tail cable 2 No 2 Use of MotorTurbo...

Страница 29: ...rovide circuit diagram bill of material legends calibration rules and other maintenance materials to other organizations 3 Should the quality assurance complaint be reasonable Changzhou Sifary Medical...

Страница 30: ...ase deal with them according to the local environmental protection laws and regulation Rights All rights of modifying the product are reserved to the manufacturer without further notice The pictures a...