11 EMC Tables
Page
30
/
35
Surge
IEC 61000-4-5
+/-1 kV
differential mode
+/-2 kV common
mode
+/-0.5 & 1 kV
differential mode
+/-0.5, 1 & 2 kV
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
100 % Dip; for
0,5 cycles
At 0
°
, 45
°
,
90
°
, 135
°
,
180
°
, 225
°
,
270
°
and 315
°
60% Dip for
1 Cycle
30% Dip for
25 Cycles
100% Dip for
5 Seconds
100 % Dip for 0,5
cycles
At 0
°
, 45
°
,
90
°
, 135
°
,
180
°
, 225
°
,
270
°
and 315
°
60% Dip for
1 Cycle
30% Dip for
25 Cycles
100% Dip for
5 Seconds
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the
HyperLight
requires
continued operation during
power mains interruptions, it
is recommended that the
HyperLight
be powered
from an uninterruptible
power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30A/m
3 & 30A/m
Power frequency magnetic
fields should be that of a
typical commercial or
hospital environment.
Guidance and Manufacturer
’
s Declaration
–
Electromagnetic Emissions
The HyperLight is intended for use in the electromagnetic environment specified below.
The customer or user of the HyperLight should ensure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance