Model: HyperLight 

 

 
 
 
 
 
 
 

Hand-held Dental X-ray System 

USER MANUAL 

Changzhou Sifary Medical Technology Co.,Ltd.

Содержание HyperLight

Страница 1: ...Model HyperLight Hand held Dental X ray System USER MANUAL Changzhou Sifary Medical Technology Co Ltd...

Страница 2: ...Version S01 IFU 7035001 Issued 2021 6 1 Size 197mmX140mm...

Страница 3: ...ng 11 6 Operation 12 6 1 Operation panel instructions 12 6 12 PowerOn Off 13 6 3 Screen instructions 14 6 4 Set interface 16 6 5 Use of exposure function 17 6 5 Positioning Instructions 18 6 5 1 Paral...

Страница 4: ...1 Overview Page 4 35 1 Overview 1 1 Content 4 3 1 2 1 HyperLight Main Body 2 Base 3 Power Adapter with removable power cord 4 Backscatter Radiation Shielding...

Страница 5: ...1 Overview Page 5 35 1 2 Packing list and coding HyperLight Main Body 1 pcs Base 1 pcs Power Adapter with removable power cord 1 pcs Backscatter Radiation Shielding option...

Страница 6: ...number Type BF applied part Alternating current Dispose of in accordance with the WEEE directive Keep dry 55 C 20 C Temperature limitation 80 20 Humidity limitation 106kPa 70kPa Atmospheric pressure...

Страница 7: ...ered justified the practitioner must take the appropriate precautions such as use of radiation safety garments to limit radiation exposure beyond the maxillofacial complex Pediatric The medical practi...

Страница 8: ...t lamps radio transmitters remote controls and do not use this system near the active HF Surgical Equipment in the hospital Portable RF communications equipment including peripherals such as antenna c...

Страница 9: ...n regulations Pregnant women should not be exposed to X rays unless necessary Proper safety precautions should be taken to minimize dose to the fetus Operators must be fully acquainted with industry s...

Страница 10: ...4 Safety Precautions Page 10 35 4 2 Leakage Dose Leakage Dose Permissive Range 65 kVp 2 5 mA 0 5 s Max Exposure Condition At Focal Spot to Distance 1 m 1 30 Duty Cycle 0 25 mGy h...

Страница 11: ...ther end into the socket The charging icon will be displayed on the screen at this time External extension cord interface Charging port NOTE Extension cords are not provided 5 2 Install backscatter ra...

Страница 12: ...tions Symbol Type Choose adult mode or child mode Choose tooth type Increase exposure time Reduce exposure time Press two buttons at the same time to lock or unlock the machine Press these two buttons...

Страница 13: ...Power On 1 First press the power button and then let go then the screen will enter the boot interface 2 Make sure that the battery level is not lower than the figure below Battery level 1 6 2 2 Power...

Страница 14: ...2 3 4 5 6 7 No Item Description 1 Bitewing Selects the tooth type Lower Molar Upper Molar Canine Incisor 2 Digital Sensor Use the sensor to receive X rays Phosphor Plate Use the Phosphor Plate to rec...

Страница 15: ...cator light starts to red glow it means the battery needs to be charged Battery Charging Indicator Indicates the battery charger is connected to the device 5 Tube Voltage Curre nt Indicator Indicates...

Страница 16: ...s at the same time to enter the setting interface press the to turn the page and press the and the to adjust the displayed parameters Display content Meaning Adjust the volume of the machine Select th...

Страница 17: ...e the screen will display the inclination angle of the machine to the ground plane and the 60S countdown 3 At this time you need to press and hold the exposure button again for exposure Warning During...

Страница 18: ...reen display is finish it means that the exposure has been completed 6 5 Positioning Instructions 6 5 1 Paralleling technique The sensor is placed in a holder which is used to align the sensor paralle...

Страница 19: ...or in place with his her finger The X ray beam is directed perpendicularly towards an imaginary line which bisects the angle between the sensor plane and the long axis of the tooth 2 1 Position the tu...

Страница 20: ...ly Exposure time s Digital Sensor Adult Bitewing 5 8 0 4 Lower Molar 5 0 2 Upper Molar 30 0 32 Canine Maxilla 45 Mandible 20 0 25 Incisor Maxilla 45 Mandible 25 0 16 Child Bitewing 5 8 0 32 Lower Mola...

Страница 21: ...30 0 25 Canine Maxilla 45 Mandible 20 0 13 Incisor Maxilla 45 Mandible 25 0 1 Film Adult Bitewing 5 8 0 8 Lower Molar 5 0 4 Upper Molar 30 0 63 Canine Maxilla 45 Mandible 20 0 5 Incisor Maxilla 45 Ma...

Страница 22: ...Radiation Shielding Do not use cleaning agents other than ethanol since certain chemical combinations may deteriorate the HyperLight plastic prematurely 2 Unplug the charging cradle before cleaning NO...

Страница 23: ...8 Maintenance Page 23 35 8 Maintenance If you do not use the machine for a long time you need to charge the machine every 6 months...

Страница 24: ...nction Causes Methods Insufficient charge Use the charger to charge the machine The machine temperature is too high Let stand for more than half an hour Abnormal exposure time During the exposure proc...

Страница 25: ...rent 2 5mA 20 Exposure time adjustment range 0 02 2S Nominal power 162 5W nominal at 65kV 2 5mA Inherent Filtration 0 8mmAl Total Filtration 1 8mmAl Duty Cycle 1 30 Minimum source to skin distance 20c...

Страница 26: ...ical Data Page 26 35 Operating altitude 3000m above sea level Transport and storage conditions Ambient temperature 20 C 55 C Relative humidity 20 80 non condensing at 40 C Atmospheric pressure 70 kPa...

Страница 27: ...minal tube voltage 70kV X ray source assembly maximum heat content 4500J Maximum anode cooling rate 110W Nominal anode input power 600W Target material Tungsten Minimum target angle 12 Filament voltag...

Страница 28: ...10 Technical Data Page 28 35...

Страница 29: ...network that supplies buildings used for domestic purposes Harmonic emissions IEC61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacturer s declaration...

Страница 30: ...quality should be that of a typical commercial or hospital environment If the user of the HyperLight requires continued operation during power mains interruptions it is recommended that the HyperLigh...

Страница 31: ...meters Field strengths from fixed transmitters as determined by an electromagnetic site survey should be less than the compliance levels V1 and E1 Interference may occur in the vicinity of equipment c...

Страница 32: ...res may be necessary such as re orienting or relocating the Model 005 2 Over the frequency range 150 kHz to 80 MHz field strengths should be less than 1 V m 3 The ISM industrial scientific and medical...

Страница 33: ...istance for the higher frequency range applies NOTE 2 The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 28...

Страница 34: ...tal Synthetic materials electrical components and printed circuit boards are disposed as electrical scrap Please deal with them according to the local environmental protection laws and regulation Righ...

Страница 35: ...Changzhou Sifary Medical Technology Co Ltd Add NO 99 Qingyang Road Xuejia County Xinbei District Changzhou City 213000 Jiangsu P R China Tel 86 0519 85962691 Fax 86 0519 85962691 Email ivy sifary com...