method has to be used, please valid-ate it
prior to use.
Warning
Use only approved washer-disinfectors
according to EN ISO 15883, maintain
and calibrate it regularly.
Follow instructions and observe
concentrations given by the
manufacturer (see general
recommendations).
Disinfection:
Automated Thermal Disinfection in
washer/disinfector under consideration of
national requirements in regards to A0
value (see EN ISO 15883). A
disinfection cycle of 5 min disinfection at
93°C has been validated for the device to
achieve an A0 value of 3000.After manual
cleaning, the instruments should be
automated disinfected of sterilized
immediately. A manual disinfection is not
recommended.
Drying:
Automated Drying:
Drying of outside of instrument through
drying cycle of washer/disinfector. If
needed, additional manual drying can be
performed through lint free towel. Insufflate
cavities of instruments by using sterile
compressed air.
Functional
Testing,
Maintenance
Visual inspection for cleanliness of the
instruments and reassembling. Functional
testing according to instructions of use. If