Manufacturer declaration – Electromagnetic noise immunity
Noise immunity test
IEC 60601 – test
level
Correspondence
level
Electromagnetic enviroment -
Guide
Portable and mobile radio transmitter should
not be use within the recommended safety
distance to the SWIFTCAM (including the
lines)!
The recommended safety distance will be
calculate by the appropriate transmitting
frequency formula.
Recommended safety distance:
Guided HF-disturbances acc. to
IEC 61000-4-3
3 V
eff
150 KHz to 80 KHz
3 V
eff
d=1,2
√
P
Radiated HF-disturbances acc. to
IEC 61000-4-3
3 V / m
80 MHz to 2,5 GHz
3 V / m
d=1,2
√
P 80 MHz to 2,5 GHz
d=2,3
√
P 800 MHz to 2,5 GHz
P – transmitter wattage rating
(W) acc. to manufacturer
(transmitter)data
d – recommended safety disdance in
meters (m)
The electrical field strength of fixed radio
transmitter is lower than the correspondence
level
b
according to a fieldwork
a
.
In the enviroment of equipment which bore the
following pictograph, it is possible to appear
interferences.
Note 1: At 80 MHz and at 800 MHz the higher result is valid.
Note 2: This guideline may not be the case for all situations. The spread of electromagnetic waves will
be affect by absorption and reflexion of buildings, objects and humans.
a
b
The electric field strength of fixed transmitter, like wireless telephones, mobile radio services, AM- and FM
broadcast transmitter and TV transmitter, couldn’t be predetermine correctly. In order to ascertain the
electromagnetic enviroment at fixed HF-transmitter, it is commendable to inspect the location. If the ascertained field
strength at the SWIFTCAM location exceed the specified correspondence level, it’s may be necessary to adopt other
means, e.g. reorientation or removal the SWIFTCAM.
By means of the frequency range from 150 kHz to 80 MHz the field strength is less than
3 V/m