MAINTENANCE
Changing the battery
1) When the battery level is low, the screen will flash slowly about once every 2 seconds.
If the flashing is continuous then the battery is below the minimum level and should be replaced.
2) Undo the back cover of the device by undoing the securing screws in the corner of the backplate.
Remove the battery from the battery holder and replace with a long life, low drain 1.5 volt alkaline
battery. Replace the backplate and tighten all the corner securing screws.
Replacing the 22 mm Adaptor seal
1) There is an O-ring seal in the 22 mm Adaptor that prevents the leakage of gas from the
anaesthetic circuit. If this seal becomes damaged or is lost it must be replaced. Use a small
screwdriver to ease out the old seal and then press the new seal into place.
Dichtung mit einem kleinen Schraubendreher heraus und drücken dann die neue Dichtung hinein.
2) New seals can be obtained from EICKEMEYER® under Replacement Part number E21319501.
TROUBLESHOOTING
Problem:
There are 6 horizontal lines on the screen all the time.
Solution:
The re-calibration procedure was executed incorrectly. Please restore the factory defaults
Problem:
The display is constantly flashing although the numbers seem correct.
Solution:
The internal AA battery is low. Replace the battery.
Problem:
There is nothing on the display and no response to the ON / OFF button.
Solution:
Fit a new Long-life AA battery. If this does not resolve the problem, contact EICKEMEYER
®
to arrange a unit service.