13. ELECTROMAGNETIC IMMUNITY
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The OxyVet IV Oxygen Concentrator is intended for use in the electromagnetic environment specifi ed below. The customer
or the user of the OxyVet IV Oxygen Concentrator should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be use no closer to any
part of the OxyVet IV Oxygen Concentrator
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = √P
( )
3.5
V
1
d = √P
( )
3.5
E
1
= 80 MHz to 800 MHz
d = √P
( )
7
E
1
= 800 MHz to 2.5 GHz
Where
p
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d
is the recommended separation
distance in metres (m).
b
Field strengths from fi xed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic is affected by absorption and refl ection from
structures, objects and people.
a
Field strengths from fi xed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey should be
considered. If the measured fi eld strength in the location in which the OxyVet IV Oxygen Concentrator is used exceeds
the applicable RF compliance level above, the OxyVet IV Oxygen Concentrator should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating
the OxyVet IV Oxygen Concentrator.
b
Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.