KGW
1- 6
Function Key to store and to delete Tare Values (TARE- Button)
•
(Subtractive) TARE function is activated.
•
Current weight value is stored into tare memory.
•
Display
shows
″
0
″
.
•
Net -LED is illuminated.
•
Exit by repressing TARE button and GROSS weight is displayed again.
Function Key for Zero-Setting (Zero- Button)
•
Sets display to
″
0
″
(semi-automatic zero-setting device).
•
Zero-setting range: -1…+3% of nominal capacity.
Zero-setting outside semi-automatic zero-setting range and also in active
tare weighing mode is
not
possible. Outside of scales 1/4d-range (a quarter
of verification interval) only +/- sign is illuminated.