PerfiLab Manual 

EGITRON

 

 

 

Page 15 

Acquiring the bottleneck profile 

To  acquire  the  bottleneck  profile,  after  having  done  steps  1  through  9  listed  above, 

initialize the software “PerfiLab Pro” and follow the steps listed below: 

 

1.  In the main window press the “Add New Test” button: 

 

 

2.  After naming the test and filling all the identification fields, save the test using 

this button:  

 

3.  To  collect  diameter  values  at  0  degrees  and  if  the  PerfiLab  is  in  its  initial 

position, press the acquisition button: 

 

4.  If the PerfiLab is not in its initial position (this information is displayed in the status bar) 

and  before  taking  step  number  3,  push  the  Initialize  button  to  make  the 

PerfiLab go to its initial position: 

 

5.  If  the  purpose  of  the  test  is  to  collect  values  at  0  and  90  degrees,  after  the  rotative 

base is rotated 90 degrees, press the acquisition button again (step number 3) to start 

the diameter reading at 90 degrees. 

 

6.  The acquired values are automatically stored in the Database. 

If the test was interrupted, either by pressing the emergency button on the command box 

frontal  panel  (in  this  case  the  system  will  beep  intermittently)  or  by  clicking  on  the  Stop 

button in the software, the user should press the Initialize button, to make the system go to 

its initial position (step number 4). 

 

Throughout the test, the collected diameter values are placed directly into the existing table and 

it’s possible to view the bottleneck profile on the graphic, as shown in figure 18.  

 

The  profiles  of  the  different  bottles  can  be  viewed  by  using  the  selection  list  placed  on  the 

bottom of the profile drawing. 

Figure 17 

Содержание PerfiLab PRF-2014-01

Страница 1: ...Interface Software for PC PerfiLab Pro v2 0 1 or above Valid for embedded Software v1 4 or above Valid for Model PRF 2014 01 or above RUA CENTRAL DA VERGADA 1280 4535 166 MOZELOS VFR PORTUGAL TEL 351...

Страница 2: ...PerfiLab Manual EGITRON II...

Страница 3: ...r equipment may be subject to change without prior notice EGITRON is not responsible for damages directly or indirectly caused by errors omissions or discrepancies between the PerfiLab and the instruc...

Страница 4: ...ld be kept to reduce damage risk to the worker and also the equipment Please read carefully the following precautions and be mindful of the warnings included throughout the manual Keep away from child...

Страница 5: ...ording to its own criteria will repair or replace without any charges with the exception of the following situations Any product and or parts that have been replaced will be EGITRON property The guara...

Страница 6: ...delivered or if possible contact your company s maintenance services If the problem is still unsolved then contact EGITRON having with you the following information Equipment name Serial number Purcha...

Страница 7: ...ons 9 2 3 Gauge configurations 12 Chapter 3 Performing the tests 13 Chapter 4 Test consult 17 Chapter 5 Printing reports 18 Chapter 6 Export data to XLS XLSX Files 19 Chapter 7 Tools 20 7 1 Calibratio...

Страница 8: ...easuring Device Rotating Accessory 4 Transversal Movement Break unavailable 5 Vertical Movement Speed Limiter 6 Supporting Head of Vertical Movement 7 Sensor for Vertical Movement Limitation 8 Measuri...

Страница 9: ...le in one single test Bottle base maximum diameter 150mm Inner bottleneck diameter 10 to 30mm Bottle length 100 to 500mm Depth control configuration between 0 and 55 mm Test resolution configuration m...

Страница 10: ...nsible for the inner measurement of the bottleneck Messages Messages are used in this manual to call for attention when important information is displayed Each type of message is identified through an...

Страница 11: ...the PC software it can also be sent by email Gauge manual and calibration certificate 1 2 Setting the mechanical system Cable connections The cables must be connected to the frontal panel of the comm...

Страница 12: ...plate found on the metallic command box rear panel matches your voltage plug 1 3 Minimum necessary requirements PC The minimum necessary requirements for the PerfiLab Pro software are o 2 GHz Pentium...

Страница 13: ...e setup exe NOTE Throughout the installation process an analysis to the PerfiLab minimum requirements will be made If a message appears requiring the Microsoft NET Framework 3 5 find a folder named Mi...

Страница 14: ...communication configuration is performed in the communication and general parameters configuration window figure 5 opened through the Communication and General Parameters option in the Configuration...

Страница 15: ...ompany name o Document code corresponding to the printed report o Header title The report configuration window figure 6 can be reached through the option Reports Header in the configuration menu To se...

Страница 16: ...rol bottles that have a step like angle at the entrance of the bottleneck When you wish to measure that step the gauge s measuring tips need to be protected during the downward movement If this is the...

Страница 17: ...s It s advised to include all the depths even at 0 mm for which you want to perform tests with a particular bottle type Test type configurations This table has the purpose of simplifying the process o...

Страница 18: ...e software main window As previously mentioned it is also possible to associate a depth control table to a bottle type and like this the table is automatically loaded when that bottle type is selected...

Страница 19: ...he measuring arm D LCD Display H MODE TOL Button I Housing The interior measuring device gauge is factory configured in a way that allows the correct system functioning with the following parameters o...

Страница 20: ...at position To perform a test follow the next steps and consult the figures 14 and 15 Placing the bottle 1 Turn on the mechanical system through the ON OFF button in the metallic command box s rear pa...

Страница 21: ...gh the transversal movement handle see figure 16 To know when the centre is reached move the handle and at the same time watch the measuring device display Stop where the maximum diameter is reached 1...

Страница 22: ...5 If the purpose of the test is to collect values at 0 and 90 degrees after the rotative base is rotated 90 degrees press the acquisition button again step number 3 to start the diameter reading at 90...

Страница 23: ...test in progress In this situation a beep will be heard then the user should press the Start button Add new bottles to the current test Remove bottles from the current test Export the data to Microsof...

Страница 24: ...day and the other By default the software displays the tests performed in the last 30 days After that select from the test list the one you wish to see Method number 2 Open the test Management window...

Страница 25: ...To print the report the Print button in the main window has to be clicked After this a new window will be opened and in this window the printing options are selected After this the user only has to cl...

Страница 26: ...en the export window like the one in figure 24 All the fields in this window must be filled The user also has the option of saving this exporting configuration as default To finally export the data si...

Страница 27: ...onfiguration menu Movements can have an upward or downward direction and the number of movements and depth of them can be selected in this window To start the movement specified the Move button has to...

Страница 28: ...ator Please note that this movement should be done manually and without exerting too much strength and at the same time it must allow a free sliding of the measuring device transportation system To ad...

Страница 29: ...nt with the certificate from the manufacturer consult the calibration chapter in the gauge manual If even after the calibration process the measure inaccuracy remains please contact our technical assi...

Страница 30: ...able for tests with measurements at zero degrees With this feature is also available a button that lets you clear the whole grid of values 8 4 Mechanical tuning This procedure should be performed by a...

Страница 31: ...ing This motor centring tuning procedure should be followed with the help of figure 29 Before starting the procedure be aware of The coupling status check if any of the spiral blades is damages if tha...

Страница 32: ...ct points where the bottleneck profiles should be tested at 0 and 90 during the tests To use it follow the next steps see figure 30 1 Place the conic marker inside the bottleneck with two marks that c...

Страница 33: ...shaft that holds the measuring head as well as the helicoidally screw motor movement both have to be regularly lubricated Following the same principle the system command box should be kept in a dry a...

Страница 34: ...emory from 2 Kbytes to 16 Kbytes RAM Interfaces One RS232 C port with a standard connector DB 9 male to connect to a personal computer One port with a standard connector DB 9 female for the connection...

Страница 35: ...N Engenharia e Automa o Industrial Lda Marque Mark Marque Produto PERFILAB Controlo Autom tico do Perfil de Gargalos de Garrafas Produit Product Producto Modelo PRF 2014 01 N mero de S rie Mod le Mode...

Страница 36: ...PerfiLab Manual EGITRON Page 29 RUA CENTRAL DA VERGADA 1280 4535 166 MOZELOS VFR PORTUGAL TEL 351 227 471 120 FAX 351 227 471 129 E mail info egitron pt WEB www egitron pt...

Отзывы: