ExtraLab Plus – Cork Extraction Force and Torque Measuring System
47 / 59
2.1.5.
Fix the “star” of the extraction set on the gripper holders (Figure 2-1B).
2.1.6.
Lower the gripper using the down
arrow key at the calibration menu until
you can center the interconnection
bar on the rod clevis . Fix the
interconnection bar with the hinged
spring pin (Figure 2-1C).
2.1.7.
Connect reading unit, power cable and load cell (Figure 2-2). Turn On the reading and
ExtraLab Plus and allow both units to stabilize during 30min.
To apply load on the load cells, use the hook wrench supplied with the calibration kit, or
using an Allen key size 4mm rotate the gear at the top of ExtraLab Plus. Slowly rotate the
lock nut in the clockwise direction or the gear in the counterclockwise direction, to raise the
measurement unit (see figure below).
2.2. Readings acquisition procedure.
2.2.1.
Assuring the reference load cell has
only the load of the necessary
attachments, TARE it.
2.2.2.
Pre-load the system with an end-of-
scale value near to 95 daN.
2.2.3.
Release the system and start the
Annex C . Figure 2- 1
Annex C . Figure 2-1
Annex C . Figure 2-2
A
B
“Star”
C
Hinged
spring pin
Interconnection bar
Rod Clevis
Содержание ExtraLab Plus
Страница 2: ......
Страница 42: ...ExtraLab Plus Cork Extraction Force and Torque Measuring System 42 59...