Side and Head Flex Shields
Left and Right Head Flex Shields and Side Flex Shields attach to the Head and Side Rail on the
Base Platform. These shields block scatter radiation from below the table at the head end of the
table (patient right side shown).
Side Flex Shields (Left and Right)
There are two Side Flex Shields, one right and one left. Each Side Flex Shield has three
components:
a.
A top attachment bar that fits over and attaches to the Head Rail. When attached to the
Base Platform, the sliding support arm can be pulled away from the table using the handle
that is on the longer arm. This will cause the Arm Board to move outward as well. The
movement is intended to facilitate radial or brachial artery access by abducting the arm
about 30°. The x-ray source can be moved under the abducted arm board in order to
visualize blood vessels in the arm.
b.
A larger arm of the Side Flex Shield is attached to the head end of the arm. The Flex Shields
have an internal stay to give it a shape that helps prevent interaction with the x-ray tube
housing and creates a center of gravity that cause the Flex Shields to passively swing under
the x-ray table.
Figure 5 Right Head Flex Shield and Side Flex Shields attached to the Head and side rail.