33
Remark:
• Input impedance: 2MΩ.
• Frequency response: 40Hz~400Hz.
• Display effective value of sine wave (mean value response).
C.
DC Current
Range
Resolution
Accuracy
Overload Protection
2mA
0.0001mA
±(0.5%+5)
0.5A. 250V fast type fuse,
Ø5x20mm
20mA
0.001mA
200mA
0.01mA
±(0.8%+5)
20A
0.001A
±(2%+10)
Un-fused
Remark:
• At 20A range: For continuous measurement ≤10 seconds and interval not less than 15
minutes.
• Measurement voltage drop: Full range at 200mV.
D. AC Current
Range
Resolution
Accuracy
Overload Protection
20mA
0.001mA
±(0.8%+10)
0.5A 250V fast type fuse,
Ø5x20mm
200mA
0.01mA
±(1.2%+10)
20A
0.001A
±(2.5%+10)
Un-fused
Remark:
• At 20 A range: For continuous measurement ≤10 seconds and interval not less than 15
minutes.
• Measurement voltage drop: Full range at 200mV.
• Frequency response: 40Hz~400Hz.
• Display effective value of sine wave (mean value response).
E.
Resistance Test
Range
Resolution
Accuracy
Overload Protection
200Ω
0.1Ω
±(0.5%+10)
250V DC
or AC ms
2kΩ
0.0001kΩ
±(0.3%+1)
20kΩ
0.001kΩ
200kΩ
0.01Ω
2MΩ
0.0001MΩ
20MΩ
0.001MΩ
±(0.5%+12)
200MΩ
0.01MΩ
±[5%(reading-1000)+10]
Remark:
• At 200MΩ range , test lead is in short circuit, and it is normal to display 1000 digits. During
measurement, subtract the 1000 digits from the reading.