5.1.7 Analogue signal options
The analogue output can be used to monitor one of the four different parameters. Refer
to
Table: Analogue signal options on page 61
.
Table 19
Analogue signal options
Option number
Description of analogue output number
0
Measured pump rotational speed. This is the factory default
setting.
1
Measured system power
2
Measured motor temperature
3
Measured controller temperature
To configure the analogue output, send the command that follows (where, the
character 'd' shows the option number given in
Table: Analogue signal options on page
Command
!
S
8
7
1
sp
d
cr
The reply is as follows:
Reply
*
S
8
7
1
sp
d
cr
The analogue output signal setting is now stored in memory of the pump.
To check which analogue output signal setting is enabled, send a query as follows:
Command
?
S
8
7
1
cr
The reply will be as follows:
Reply
=
S
8
7
1
sp
d
cr
5.1.8 Factory settings
The pump can be configured again to its original factory settings with one serial
command.
To reset the controller to factory settings, send the command that follows:
Command
!
S
8
6
7
sp
1
cr
The reply will be as follows:
Reply
*
S
8
6
7
sp
r
cr
The factory settings are restored in the memory of the pump.
5.1.9 Assigning a multi-drop address
Each individual pump must be programmed with its own multi-drop address through a
point
‑
to
‑
point connection before introduction into a multi-drop network.
When the pump is shipped, multi-drop mode is disabled by default.
Page 61
B85200880_D - Configuration