8. Fault finding
Table 32
Fault finding
Symptom
The controller LEDs do not flash for 0.5 seconds when the system is set to on
on page 90
The pump does not rotate after a parallel start command is sent
The pump does not rotate after a serial start command is sent
on page 90
The pump does not respond in multi- drop mode
on page 91
on page 91
Ultimate pressure cannot be reached
on page 91
The pump is very noisy or there is excessive vibration or both
on page 92
on page 92
Fail signal or standby signal not working
on page 93
Yellow service LED is flashing a repeated sequence
on page 93
on page 93
on page 93
Fault
The controller LEDs do not flash for 0.5 seconds when the system is set to on
Cause
The electrical supply has failed.
Remedy
Make sure that the electrical supply is set to on and the fuses (and current limiting
devices) are not tripped.
Cause
The pump rotor is rotating.
Remedy
The LEDs can possibly not flash if the pump is in operation.
Fault
The pump does not rotate after a parallel start command is sent
Cause
Check that the electricity supply is on. Check if the fail output is active.
Remedy
If there is a fail signal, check if the red alarm LED is flashing. If yes, refer to
on page 94. If the power is supplied and fail signal is not given, but the
rotor does not rotate, the fault is in the pump.
Fault
The pump does not rotate after a serial start command is sent
Cause
Check if the pump returns a reply to the start command.
Remedy
If there is no reply, do the checks given in
on page 92 or do the
checks given in
The pump does not rotate after a parallel start command is sent
page 90.
07/2021 - ©Edwards Limited
Page 90
B8G000880_E