The standby speed is now stored in the memory of the pump.
To check what Standby Speed is set, send a query as follow:
Command
?
S
8
5
7
cr
The reply is as follows:
Reply
=
S
8
5
7
sp
d
d
d
cr
5.1.7 Normal speed setting
The normal speed is a user-configurable setting and can be set to the value between
50% and 100% full rotational speed. When the pump is shipped, it is configured with a
normal speed of 80% full rotational speed.
To change the normal speed setting, send the command that follows (characters 'd'
shows the value as a percentage of full rotational speed):
Command
!
S
8
5
6
sp
d
d
d
cr
The reply is as follows:
Reply
*
S
8
5
6
sp
r
cr
The normal speed is now stored in the memory of the pump.
To check what normal speed is set, send a query as follows:
Command
?
S
8
5
6
cr
The reply is as follows:
Reply
=
S
8
5
6
sp
d
d
d
cr
5.1.8 Timer setting and options
The timeout period is a user-configurable option and can be set to the value from
1 to 30 minutes.
When the pump is shipped, it is configured with a default timeout period of 8 minutes.
Refer to
on page 13 for a full description of timer functionality.
To change the timer setting, send the command that follows (where the characters 'd'
represent the timeout period in minutes):
Command
!
S
8
5
4
sp
d
d
cr
The reply will be as follows:
Reply
*
S
8
5
4
sp
r
cr
The Timer setting is now stored in the memory of the pump.
To check what time
‑
out period is set, send a query as follows:
Command
?
S
8
5
4
cr
The reply will be as follows:
Reply
=
S
8
5
4
sp
d
d
cr
07/2021 - ©Edwards Limited
Page 66
B8G000880_E
B8G000880_E - Configuration
Содержание nEXT 55
Страница 110: ...This page has been intentionally left blank...
Страница 114: ...This page has been intentionally left blank...
Страница 115: ...This page has been intentionally left blank...
Страница 116: ...edwardsvacuum com...