INSTALLATION

D059-14-880 Issue D

3

INSTALLATION

3.1

Unpacking and inspection

Remove all packing materials and protective covers and check the IS16K for damage.  If the IS16K is damaged, notify
your supplier and the carrier in writing within three days; state the Item Number of the IS16K together with your
order number and your supplier's invoice number. Retain all packing materials for inspection. Do not use the IS16K.    

3.2

General

The vacuum switch should be mounted as close as possible to the point at which the pressure is to be measured.  Use
a short branch tube with an internal diameter no less than that of the switch vacuum connection.  Long, narrow and
angled connections can cause significant errors.

This equipment is not protected from dripping water from above, unless mounted with the vacuum flange pointing
downwards.

3.3

Connection to the Vacuum System

The vacuum switch may be mounted in any orientation but mounting with the vacuum switch vertical will prevent
the possibility of debris falling into the vacuum port and affecting the performance of the switch.

3.4

Connection to the Electrical Circuit

CAUTION

Ensure that the circuit connected to the vacuum switch is suitably fused to prevent the switch contact rating 
being exceeded.  The fuse rating must not exceed 10 amps.

WARNING

Do not allow the internal pressure to be higher than 1 bar gauge (2 bar absolute). If the pressure 
in your system can be higher than 1 bar gauge, you must fit a suitable pressure relief valve. 
Overpressure of the switch could cause injury to persons.

WARNING

Ensure that the electrical installation of the Vacuum Switch conforms with your local and national 
safety requirements. It must be connected to a suitably fused and protected electrical supply and 
a suitable earth point.

WARNING

If, in your application, failure of this switch could lead to  a safety hazard, it is recommended that 
the switch should be wired in series with another interlocking device, for example chamber lid 
microswitch, to provide a double level of protection.

Содержание IS16K

Страница 1: ...D059 14 880 Issue D Original Instruction Manual Vacuum Interlock Switch IS16K Description Item Number Vacuum Interlock Switch IS16K D059 14 000...

Страница 2: ...declaration relates is in conformity with the following standard s or other normative document s EN61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use...

Страница 3: ...Mechanical Components X O O O O O O GB T 26572 O Indicates that the hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in GB T 26572 X GB T2...

Страница 4: ...This page has been intentionally left blank...

Страница 5: ...ting conditions 2 2 2 Performance 2 2 3 Mechanical data 2 2 4 Electrical data 2 2 5 Product item number 3 3 INSTALLATION 4 3 1 Unpacking and inspection 4 3 2 General 4 3 3 Connection to the Vacuum Sys...

Страница 6: ...d 2009 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited Contents Illustrations Figure Page 1 Vacuum switch plug connections 5 2 Dimensions of VS16K vacuum switch 7 Ta...

Страница 7: ...ssociated equipment and process 1 2 General description The Edwards IS16K Vacuum Interlock Switch is designed for applications where the electrical circuits within the vacuum chamber must be de energi...

Страница 8: ...664A Pollution category IEC1010 2 Enclosure rating IP40 Set point Within the range 250 to 500 mbar below atmospheric pressure Maximum switch point differential 100 mbar Leak rate 1 x 10 9 mbar litre s...

Страница 9: ...System The vacuum switch may be mounted in any orientation but mounting with the vacuum switch vertical will prevent the possibility of debris falling into the vacuum port and affecting the performanc...

Страница 10: ...land strain relief is effective a cable with a minimum diameter of 6 mm must be used Select a three or four core cable of a rating suitable for the power supply to be switched One core should be colou...

Страница 11: ...re trademarks of Edwards Limited OPERATION D059 14 880 Issue D 4 OPERATION 4 1 Adjustment The IS16K Vacuum interlock switch is pre set to operate in the range of 250 to 500 mbar below atmospheric pres...

Страница 12: ...nd the pin marked Lo Refer to Figure 1 for pin connections 3 With the IS16K at atmosphere verify that the meter reads open circuit 4 Reduce the pressure to below 500 mbar as measured on a vacuum gauge...

Страница 13: ...ND DISPOSAL D059 14 880 Issue D 6 STORAGE AND DISPOSAL 6 1 Storage Store in a cool dry place Do not exceed the limits specified in Section 2 6 2 Disposal Dispose of the IS16K in accordance with any lo...

Страница 14: ...D059 14 880 Issue D Page 8 Edwards Limited 2009 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Страница 15: ...of Vacuum equipment and Components Procedure HS1 If you are returning a vacuum pump note the following If a pump is configured to suit the application make a record of the configuration before return...

Страница 16: ...P800 80 000 Issue T edwardsvacuum com...

Отзывы: