D0300030001EN - Technical data
Page 4
Technical data
General gauge head data
Table 1 General gauge head data
Description
IG40 BA
IG40 EX
Materials
Lead-in pins
NiFe
Insulator
Al
2
0
3
ceramic
Pin sealing plate
NiFe
Flange
Stainless steel
Cathode
Iridium with yttrium oxide coating
Anode
Pt/Ir 90/10 and Mo/Pt
wrapped wire
Stainless steel 1.4404
Collector
Tungsten
Reflector
NiFe
Flange connection
DN 40 CF
Electrode system configuration
Bayard-Alpert
Extractor system
Ambient temperature during
operation
20 to 80 °C
Maximum flange temperature with
bakeable gauge cable
250 °C
Maximum bakeout temperature
(with no cable connected)
400 °C
Maximum recommended rate of
temperature change during bakeout
5 °C per minute
Storage temperature
20 to 50 °C
Relative humidity
Annual average
65% (non-condensing)
On 60 days
85% (non-condensing)
Use
Indoors (altitude 2000 m maximum above sea level)
X-ray limit
< 1 x 10
-11
mbar (hPa)
< 1 x 10
-12
mbar (hPa)
Upper limit of measuring range
1 x 10
-2
mbar (hPa)
1 x 10
-4
mbar (hPa)
Lower limit of measuring range
2 x 10
-11
mbar (hPa)
2 x 10
-12
mbar (hPa)