Operation
4.3
Bakeout band control (EXC120/300/300M only)
Note:
As described in
, you must insert a suitably rated fuse into the bakeout band fuse holder before
you can use a bakeout band.
Press the Heater switch on the front panel of the Controller (Figures
and
, item 2) to select the bakeout band on.
The Controller will switch on the electrical supply to the bakeout band when the EXT pump speed has reached the
Normal speed setpoint. The electrical supply to the bakeout band will switch off if the pump speed drops to below
the Normal speed setpoint, if the Controller is switched off, or if the Controller trips into Fail condition or Axial
Emergency condition.
The LED on the Heater switch is on when the electrical supply to the bakeout band is on.
4.4
Operation with high inlet pressure
If the EXT pump inlet pressure rises, the power supplied by the Controller to the pump-motor will increase to
counteract the gas frictional load. The pump rotational speed will remain constant until the Controller peak power
level is reached; beyond this power level, the speed of the pump will start to reduce.
If the pump speed falls to below 50% of its full rotational speed, the Controller may trip into Fail condition, depending
on how you configured the Controller (see Sections
and
).
Refer to the EXT pump instruction manual for the maximum allowable inlet pressure.
4.5
Operation with high pump temperature
Temperature sensors in the Controller and the EXT pump are monitored by the Controller. If the Controller detects
that the pump temperature is too high, the power supplied to the pump-motor is reduced; the pump may not
therefore be able to maintain full rotational speed if it is too hot.
If the pump speed falls to below 50% of its full rotational speed, the Controller may trip into Fail condition, depending
on how you configured the Controller (see Sections
and
).
Refer to the EXT pump instruction manual for the pump operating temperature ranges.