Installation
D396-14-880 Issue J
Figure 12 - Top cover of the Controller
3.13
Configure the Controller
3.13.1
Introduction
The Controller has four DIP switches which can be used to configure the Controller for your application. The DIP
switches are on the top of the Controller (the CONFIG switches shown in
). Set the DIP switches as described
in the following sections.
3.13.2
Enable/disable the Internal Timer to monitor low pump speed
Notes:
1. If you pump a high gas load with the Internal Timer enabled (DIP switch 1 set to the off position), the
EXT pump may stall before the Controller trips into Fail condition. Ensure that oil which backstreams
from the backing pipeline will not adversely affect your process.
2. Disable the Internal Timer to provide the greatest protection against backstreaming when Fail condition
occurs; as supplied, the Internal Timer is enabled.
Set DIP switch 1 to enable or disable the use of the Internal Timer when the pump rotational speed falls to below 50%
of full rotational speed during operation (see
):
z
Set DIP switch 1 to 'on' to disable the timer.
The Controller will then trip into Fail condition as soon as the
pump rotational speed falls to below 50% of full rotational speed.
z
Set DIP switch 1 to 'off' to enable the timer.
The Internal Timer will then start as soon as the pump
rotational speed falls to below 50% of full rotational speed. If the pump speed remains below 50% of full
rotational speed after the preset time, the Controller will trip into Fail condition.