D13510840D - Interface description
Page 88
Arrays
Read single elements: Array index in first DATA-byte
Write single elements: Array index in first DATA byte and values in following DATA
bytes
Read all elements: Pseudo array index 255 in first DATA byte
Write all elements: Pseudo array index 255 in first DATA byte and values in following
DATA bytes
Response from slave (in case data are sent): Array index or pseudo array index in first
DATA byte and values in following DATA bytes
All elements of an array have the same Min/Def/Max value.
Commands
Command
dec
Command
hex
Name
Data type Read/
Write
Min-, Def-,
Max-value Meaning
0
0000
NOP
NO_DATA R/W
No operation,
replies without
data
1
0001
Start
NO_DATA W
Switch to measure
2
0002
Stop
NO_DATA W
Switch to standby
3
0003
Vent
NO_DATA W
Switch to vent
4
0004
Calibration
NO_DATA W
Start,
acknowledge
calibration
5
0005
Clear error
NO_DATA W
Clear Error or
Warning
6
0006
Zero
UINT8
R/W
0 = Zero "Off"
1 = Zero "On" /
"Update"
without data =
toggle Zero state
9
0009
Emission
nominal
status
UINT8
R/W
Emission nominal
status
0 = OFF
1 = STANDBY
2 = ON
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......