D13510840D - Interface description
Page 76
ASCII Protocol
Communication Parameters
Preface
The RS232 journal uses an ASCII format: so every leak detector can be controlled with a
simple terminal program.
Choose the interface journal "ASCII" to approach the leak detector with the computer via
RS232 interface.
In case the leak detector shall also be controlled via computer (START, STOP, ZERO, etc.), affix
the location of control at the leak detector to "All", "RS232" or "Local and RS232".
Interface connecting cable
The interface (RS232) is wired as data communication equipment (DCE). The connection is
made via a 9-way sub-D socket at the leak detector. The signals are assigned as follows:
The other pins are not used.
The levels on the RS 232 interface are defined as follows:
A standard RS232 cable can be used (straight-through connecting cable, RxD and TxD not
crossed). The RS232 hardware handshake must be switched off (in RS232 control program
written by the user).
Data format
Baudrate
19200 baud, 8 data bits, 1 stop bit, no parity.
Note:
The command list in this interface description is valid from software version 3.0. Older
software versions may not contain all functions of the command list.
Pin
Name
Signal
2
RXD
Receive data (ELD500 --> PC)
3
TXD
Transmit data (PC --> ELD500)
5
GND
Reference ground
Level
Low (L)
High (H)
Voltage range
-3 V to -25 V
3 V to 25 V
Logic state
logical 1
logical 0
Level designation
Mark
Space
Содержание ELD500 Series
Страница 9: ...D13510840D Contents Page vii...
Страница 10: ...This page has been intentionally left blank D13510840D Contents Page viii...
Страница 12: ...D13510840D Introduction Page 2 This page has been intentionally left blank...
Страница 38: ...D13510840D Page 28 This page has been intentionally left blank...
Страница 58: ...D13510840D SL Sniffer line Page 48 This page has been intentionally left blank...
Страница 64: ...D13510840D Partial flow adaptor Page 54 This page has been intentionally left blank...
Страница 78: ...D13510840D Calibrated leak He 4 6 Page 68 This page has been intentionally left blank...
Страница 84: ...D13510840D Calibrated leak screwed socket Page 74 This page has been intentionally left blank...
Страница 121: ......