© Edwards Limited 2016. All rights reserved.
Page 35
Edwards and the Edwards logo are trademarks of Edwards Limited.
Operation
D397-01-880 Issue E
Figure 25 - Controlled item
2. Select the controlling item. The top highlighted line is used to select the controlling item. Select the controlling
item from the list of those available and press the 'Select' (
) button to confirm the choice.
3. Enter the required setpoints. The 'On' and 'Off' setpoints can be adjusted to suit the application. If the
controlling item is a gauge, the units can be changed between pressure (current pressure units) and voltage (V).
Select the units to be used and then adjust the 'On' and 'Off' setpoints as required as shown in
Figure 26 - Control setup
Note:
For gauges: Off setpoint
≥
On setpoint.
If the adjustment of either the On or Off setpoint results in the above rules being broken, the setpoint that
was not adjusted will be altered to match the newly entered one.
Note:
When entering a pressure, the sign of the exponent can only be changed when the exponent is non-zero. To
set a negative exponent, the exponent value should be set first, and then the 'Menu' (
) button used to
move back to alter the sign. See
.
4. Enable the setpoint. Once configured, the setpoint should be enabled by changing the bottom 'Setpoint' line
from 'OFF' to 'ENABLED'.
4.12
Parameters/Units
This screen allows the user to change the units that are displayed and other parameters such as:
Setup lock - When the 3 digit lock code is entered, the lock is enabled and an operator will not be able to
change any of the setups, however the operator is still able to scroll through the menus and start and stop
pumps. The lock is disabled by entering the 3 digit unlock code. Lock - 501, unlock - 147.
Panel Lock - This function completely locks the front panel. An operator will only be able to see the view
screen. Lock - 509, unlock - 824.
The 0 - 10 V analogue output on the logic interface can be set to follow any gauge pressure.
gea/d39701/f2
gea/d39701/
Содержание D397-01-000
Страница 2: ...This page has been intentionally left blank ...
Страница 53: ...This page has been intentionally left blank ...
Страница 54: ...This page has been intentionally left blank ...
Страница 56: ...P800 80 000 Issue T edwardsvacuum com ...