Technical data
D397-01-880 Issue E
2.4.4
Analogue outputs
Figure 4 - Pin connection for 4-way analogue output connector
Connector type
Phoenix 2.5 mm 4-way
mating half Phoenix FK-MC 0.5/4-ST-2.5
(available from Edwards as 5000 17005)
Short circuit protection
All signals
Max load
Gauge signals
±
1 mA
Common
±
3 mA
Output voltages
Gauge connected
Gauge signal voltage (
±
2%
±
15 mv when in range 0 to +10 V)
Gauge disconnected
12 V (
±
1.3 V)
Table 5 - Pin allocation for analogue outputs
Pin
Allocation
Gauges 1-3 (Top Ports)
Gauges 4-6 (Lower Ports)
1
Gauge 1 signal
Gauge 4 signal
2
Gauge 2 signal
Gauge 5 signal
3
Gauge 3 signal
Gauge 6 signal
4
Common (0 V)
Common (0 V)