© Edwards Limited 2009. All rights reserved. 

Page 11

Edwards and the Edwards logo are trademarks of Edwards Limited.

OPERATION

D141-32-880 Issue A

Table 3 - Using the GasCheck G3 functions

Image

Function
Contrast
 adjusts the contrast of the GasCheck G3 LCD display. Significant variations in 
ambient temperature conditions may cause the display to appear too dark or feint. 
When selected a number appears indicating the contrast as a percentage; use the
and        keys to adjust the display contrast to the desired level. Press the           key 
to exit to the main menu.
Backlight selects whether the backlight is on or off. In daylight the user may not be 
able to determine if the backlight is on or off therefore a symbol on the main running 
screen shows the status. Use the        or        key to move the 'tick' to the desired 
position and then press the         key to return to the Main Menu.

Note:

The backlight significantly reduces battery life.

Sound switches the audible indication of the leak rate on and off. The frequency of 
the signal increases as detected gas levels increase. Use the        or        key to move 
the 'tick' to the desired position and then press the         key to return to the Main 
Menu.

Note:

An audible 'beep' can also be heard when ever the key pad is pressed, this beep 
cannot be switched off.

Peak hold can be used to record the highest detected reading on the display until the        
key is pressed. Use the       or        key to move the 'tick' to the desired position and 
then press the         key to return to the Main Menu. When using the GasCheck G3 in 
peak hold mode, press the         key to clear the held reading. The reading held on 
screen will be logged when the         key is pressed.
Calibration can be used to switch between the factory calibration setting and user 
defined calibration parameters. Factory calibration cannot be adjusted. User defined 
calibration is described in 

Section 5.2

Battery selection can be used to select whether the batteries fitted are Alkaline (non-
rechargable) or Nickel Metal Hydride (rechargeable). Select the battery type using the        
or         key to move the 'tick' to the desired position and then press the          key to 
return to the Main Menu.

Note:

Failing to select the correct battery type will not damage the instrument or 
affect it's ability to detect gas, however it will cause the battery indicator to 
read incorrectly.

View data allows 10 readings to be logged in internal memory and recalled later. To 
store a reading press the          key while on the main running screen. When readings 
are stored in memory a symbol will appear on the Main running screen. When the 
memory is full the '    ' symbol will flash. The stored data can be viewed by scrolling 
through the logged readings using the         and        keys. To delete the stored data 
press and hold the          key, the '     ' symbol will flash, continue holding the         key 
until the data is deleted. Press the          key to exit to the Main menu.
Units selects the leak rate units displayed on the LCD screen. Select the required units 
using the        or        key to move the 'tick' to the desired position and then press the        
key to return to the Main Menu. The following units are available:

cc/sec= Cubic Centimetres per second (volumetric leak rate)
ppm= Parts per million, (concentration) 
mg/m3 = Milligrams per metre cubed (concentration)
g/yr= Grams per year (an alternative volumetric leak rate)

Содержание D141-32-000

Страница 1: ...D141 32 880 Issue A Original Instruction Manual GasCheck G3 Description Item Number GasCheck G3 D141 32 000...

Страница 2: ...boratory Use General Requirements EN61326 1 2013 Electrical equipment for measurement control and laboratory Class B Emissions Use EMC requirements General requirements Basic Immunity EN50581 2012 Tec...

Страница 3: ...Getting started 7 4 2 GasCheck G3 running screen 8 4 3 Using the GasCheck G3 8 4 4 GasCheck G3 keyboard 9 4 5 Probe options 12 5 MAINTENANCE 14 5 1 Replacing fitting batteries 14 5 2 Calibration 14 5...

Страница 4: ...tartup 7 5 Running screen 8 6 GasCheck G3 keypad layout 9 7 GasCheck G3 menu structure 10 8 Remove the probe cover 12 9 Change between short and long probes 13 10 Replacing batteries 14 Tables Table P...

Страница 5: ...y information is highlighted as WARNING and CAUTION instructions you must obey these instructions The use of WARNINGS and CAUTIONS is defined below CAUTION Cautions are given where failure to observe...

Страница 6: ...igital display 6 and are also given by a front mounted LED flasher 5 and as audible clicks from a rear panel loudspeaker 2 The GasCheck G3 has a convenient storage and carrying case together with a sp...

Страница 7: ...ayed as leak rates 1 4 Applications You can use the GasCheck G3 on many types of applications including those listed below z Quality assurance testing on manufactured component seals z Laboratory appl...

Страница 8: ...g temperature range 0 to 50 C Storage temperature range 25 to 70 C Materials of construction GasCheck G3 Storage case Polypropylene Polyurethane Battery requirements MN1500 or equivalent i e AA LR6 4...

Страница 9: ...ich can be used as a substitute for gases not listed in Table 1 group 5 being the highest sensitivity through to group 1 being a low sensitivity Contact Edwards for advice on which groups to use Respo...

Страница 10: ...case of the GasCheck G3 together with your order number and your supplier s invoice number Do not use the GasCheck G3 if any item is damaged Refer to Figure 3 and check that the storage case contains...

Страница 11: ...r of screens are displayed before the unit is ready to detect leaks Refer to Figure 4 Figure 4 Screens displayed during GasCheck G3 startup WARNING Do not use the GasCheck G3 in safety critical applic...

Страница 12: ...when the GasCheck G3 is switched on Failure to do so will result in inaccurate zero readings CAUTION Do not touch the probe or brass sensor housing Any changes in heat can result in significant chang...

Страница 13: ...or keys Note Variation in temperature humidity or background gas may result in a constant offset being detected on the GasCheck G3 To reset the zero hold the GasCheck G3 away from the source of conta...

Страница 14: ...D141 32 880 Issue A Page 10 Edwards Limited 2009 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited OPERATION Figure 7 GasCheck G3 menu structure...

Страница 15: ...the key to clear the held reading The reading held on screen will be logged when the key is pressed Calibration can be used to switch between the factory calibration setting and user defined calibrat...

Страница 16: ...the short probe B Figure 8 Remove the probe cover Some applications may require a longer probe in order to reach inaccessible areas The GasCheck G3 is supplied with a 300mm Long probe in the storage c...

Страница 17: ...e short probe A as shown in Figure 9 B Unscrew the probe counter clockwise Remove the spanner and the probe assembly C To replace the short probe with the long probe perform the same sequence in rever...

Страница 18: ...with the correct battery selection alkaline or nickel metal hydride as described in Section 4 4 This ensures that the battery status indicator can correctly identify the remaining capacity of the bat...

Страница 19: ...cedure Also remove the outer grey probe cover from the GasCheck G3 Select the calibration procedure on the main menu refer to Section 4 4 Select custom cal by using the or key to move the tick to icon...

Страница 20: ...erence leak Contact Edwards for more details Table 5 Calibration of the GasCheck G3 cc sec leak rate units Ensure the pressure on the portable reference leak is set correctly Select the calibration pr...

Страница 21: ...d in Section 2 1 Unpack as described in Section 3 6 2 Disposal Dispose of the GasCheck G3 components and used batteries safely in accordance with all local and national environmental safety requiremen...

Страница 22: ...ervice Engineers who have undergone comprehensive Edwards training courses Order spare parts and accessories from the nearest Edwards company or distributor When ordering state for each part required...

Страница 23: ...sec leak rate units 16 Custom calibration concentration units 15 G GasCheck G3 keypad 9 GasCheck G3 running screen 8 GasCheck G3 spares and accessories 18 Getting started 7 M MAINTENANCE 14 P Performa...

Страница 24: ...D373 77 880 Issue A Page 20 Edwards Limited 2009 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank...

Отзывы: