Страница 1: ...edwardsvacuum com CPG35K AND CP35KB PASSIVE PENNING GAUGE INSTRUCTION MANUAL Publication number D0300010001EN Issue A Original Instructions ...
Страница 2: ...3000105 CPG cable 50 m to which this declaration relates is in conformity with the following standard s or other normative document s EN61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use General Requirements EN61326 1 2013 Class B Industrial Electrical equipment for measurement control and laboratory Use EMC requirements General requirements and fu...
Страница 3: ...This page intentionally blank ...
Страница 4: ...Copyright notice Edwards Limited 2017 All rights reserved Trademark credits Edwards and the Edwards logo are trademarks of Edwards Limited D0300010001EN Copyright notice Page 2 ...
Страница 5: ...n 5 Supplied equipment 5 Unpacking and inspecting 5 Vacuum connection 5 Electrical connection 5 Maintenance 6 Cleaning of the gauge head 6 Mechanical maintenance 7 Removing and replacing the anode ring and the cathode plate 7 Detaching the permanent magnet 8 Spare parts and accessories 9 Spares 9 Accessories 9 Storage 10 Returning the product 11 Disposal 12 Separating the components 12 Contaminate...
Страница 6: ...D0300010001EN Contents Page ii List of Figures Dimensions 4 Sectional view of the CPG35K B gauge head 7 ...
Страница 7: ...ary precautions for all work to be performed and consider the safety instructions in this document Ensure that all vacuum components have not been contaminated before beginning any work If so adhere to the relevant regulations and take the necessary precautions when handling contaminated parts Communicate the safety instructions to all other users WARNING Warnings are given where failure to observ...
Страница 8: ...r or third persons Do not observe the information given in this document Do not use the product as intended Modify the product in any way conversions repair work etc Operate the product with accessories not listed in the corresponding product documentation Subject to technical alterations without prior notice The figures are not binding ...
Страница 9: ...t nameplate Validity This document applies to the following part numbers Intended use The CPG may only be used for the measurements of total pressures in vacuum systems and only when converted to a PGC201 vacuum gauge controller instrument Product Description Item number CPG35K NW40 Passive Penning Gauge D03000100 CPG35K DN40CF Passive Penning Gauge D03000110 CPG35K NW25 Passive Penning Gauge D030...
Страница 10: ...Ti Ni NiFe NiCr glass and ceramics Disruption pressure burst pressure absolute 10 bar Ambient temperature CPG35K 80 C CPG35KB 200 C Measurement volume CPG35K NW25 and NW40 approximately 18 cm3 CPG35K and KB DN40CF approximately 21 cm3 Connection flange CPG35K NW25 CPG35K NW40 CPG35K DN40CF CPG35KB DN40CF Weight CPG35K NW25 and NW40 approximately 0 30 kg CPG35K and KB DN40CF approximately 0 60 kg ...
Страница 11: ...ges All gauges should be located as close as possible to the point where the pressure is to be measured with the exception of cases where there is potential for considerable contamination of the gauge head In these cases the gauge head should be mounted a short distance away behind an elbow In the case of long and narrow connections the measured pressure will be too low mainly because of gas consu...
Страница 12: ... removing the emery paper residues and reassembly of the components any fingerprints and other residues have to be removed by rinsing the gauge head with a solvent followed by a drying period For thorough cleaning of the gauge head cell it is recommended to dismantle the permanent magnets Only then it is possible to flush out any iron filings from the measuring chamber that have been attracted by ...
Страница 13: ...uge head 3 After having inserted the anode ring it should be strictly observed that the two open sides of the anode ring are equally distant from the cathode plate and that the ignition pins don t touch the chamber distance approximately 1 mm Note Do not allow the ceramic bushing to get wet as it is able to absorb water The leakage current which results then causes the pressure to be shown on the ...
Страница 14: ...taching the permanent magnet 1 Detach the gauge head cable 2 Unscrew and remove the nut on the protection cap 3 Remove the protection cap and the magnet 4 Clean as described in Cleaning of the gauge head 5 Reassemble in the reverse order ...
Страница 15: ... on the product nameplate Description and ordering number according to the spare parts list Accessories Description Item number Spare cathode plates and discs D03000109 Spare Anode Ring D03000119 Description Item number CPG Cable 5 m D03000101 CPG Cable 10 m D03000102 CPG Cable 20 m D03000103 CPG Cable 30 m D03000104 CPG Cable 50 m D03000105 ...
Страница 16: ...priate storage leads to an increase of the desorption rate and or may result in mechanical damage of the product Cover the vacuum ports of the product with protective lids or grease free aluminum foil Do not exceed the admissible storage temperature range ...
Страница 17: ... declaration of contamination are returned to the sender at his own expense WARNING Products returned to Edwards for service or repair should if possible be free of harmful substances for example radioactive toxic caustic or microbiological Otherwise the type of contamination must be declared Adhere to the forwarding regulations of all involved countries and forwarding companies and enclose a comp...
Страница 18: ...ipment 2012 19 EU Edwards offers customers a recycling service for the product cables associated gauge heads at the end of the product s life Contact Edwards for advice on how to return the CPG and or cables for recycling WARNING Contaminated parts can be detrimental to health and environment Before beginning to work find out whether any parts are contaminated Adhere to the relevant regulations an...
Страница 19: ...n of Vacuum equipment and Components Procedure HS1 If you are returning a vacuum pump note the following If a pump is configured to suit the application make a record of the configuration before returning the pump All replacement pumps will be supplied with default factory settings Do not return a pump with accessories fitted Remove all accessories and retain them for future use The instruction in...