C417-60-880

Issue A Original

Instruction Manual

IPVA10EK Air Admittance Valve

Description

Item Number

Normally open valve - 230 V a.c.

C417-60-000

Normally open valve - 110 V a.c.

C417-60-100

Normally open valve - 24 V d.c.

C417-60-200

Normally closed valve - 230 V a.c.

C417-70-000

Normally closed valve - 110 V a.c.

C417-70-100

Normally closed valve - 24 V d.c.

C417-70-200

Содержание C417-60-000

Страница 1: ...Description Item Number Normally open valve 230 V a c C417 60 000 Normally open valve 110 V a c C417 60 100 Normally open valve 24 V d c C417 60 200 Normally closed valve 230 V a c C417 70 000 Normally closed valve 110 V a c C417 70 100 Normally closed valve 24 V d c C417 70 200 ...

Страница 2: ... Measurement Control and Laboratory Use General Requirements EN50581 2012 Technical Documentation for the Assessment of Electrical and Electronic Products with respect to the Restriction of Hazardous Substances and fulfils all the relevant provisions of 2006 95 EC Low Voltage Directive 2011 65 EU Restriction of Certain Hazardous Substances RoHS Directive i e The product s contain less than 0 1wt f...

Страница 3: ...mative document s None and fulfils all the relevant provisions of 2006 95 EC Low Voltage Directive 2011 65 EU Restriction of Certain Hazardous Substances RoHS Directive i e The product s contain less than 0 1wt for hexavalent chromium lead mercury PBB and PBDE 0 01wt for cadmium in homogeneous materials subject to the exemptions allowed by the Directive The RoHS Directive does not legally apply to...

Страница 4: ...This page intentionally blank ...

Страница 5: ...cal data 2 2 1 General 2 2 2 Performance 2 2 3 Electrical data 2 2 4 Materials 3 2 5 Item numbers 3 3 Installation 5 3 1 Unpack and inspect 5 3 2 Install the valve 5 3 3 Electrical connections 5 4 Operation 7 5 Maintenance 8 6 Storage and disposal 9 6 1 Storage 9 6 2 Disposal 9 For return of equipment complete the HS Forms at the end of this manual Illustrations Figure Page 1 Valve dimensions 4 2 ...

Страница 6: ...This page has been intentionally left blank C417 60 880 Issue A Page ii Edwards Limited 2013 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited ...

Страница 7: ... be found on this product The units used throughout this manual conform to the SI international system of units of measurement 1 2 Description The IPVA10EK valve is solenoid operated and is designed for the automatic admittance of air or vent gas into a vacuum system The vacuum system is isolated from atmosphere or the vent gas supply by a fluoroelastomer pad on the base of the valve plunger which...

Страница 8: ...voltage category II Leak rate 1 x 10 6 mbar ls 1 1 x 10 4 Pa ls 1 Venting rate 10 l in 12 seconds Maximum cycle frequency 100 min 1 Mean time to failure MTTF 500 000 cycles Response time 20 ms to open 30 ms to close Maximum pressure d c electrical supply 1 8 bar absolute 1 8 x 105 Pa a c electrical supply 1 8 bar absolute 1 8 x 105 Pa Electrical supply 24 V d c 110 V a c 1 phase 50 60Hz 230 V a c ...

Страница 9: ...ycle 100 continuous rating Protection class IP65 Electrical supply cable outer diameter 6 to 7 mm Coil insulation Class H Materials exposed to vacuum Normally open variants have an additional carbon loaded PTFE slide ring within the vacuum envelope Body Aluminium Valve seals Fluoroelastomer Actuator Stainless steel Shading rings Silver IPVA10EK normally open valves IPVA10EK normally closed valves ...

Страница 10: ...ata Figure 1 Valve dimensions IPVA10EK normally open valves IPVA10EK normally closed valves Voltage Item number Dimensions Voltage Item number Dimensions 230 V a c C417 60 000 Figure 1A 230 V a c C417 70 000 Figure 1A 110 V a c C417 60 100 Figure 1A 110 V a c C417 70 100 Figure 1A 24 V d c C417 60 200 Figure 1A 24 V d c C417 70 200 Figure 1B ...

Страница 11: ...the atmosphere Edwards recommend that a filter is fitted to the inlet of the valve to prevent contamination of the valve through the ingress of dust 3 3 Electrical connections Refer to Figure 2 for item numbers in brackets and use the following procedure to connect the electrical supply to the valve ensure that a suitably sized electrical supply cable is used refer to Section 2 3 1 Ensure the elec...

Страница 12: ...igure 2 5 Push the connector block 2 back into the electrical supply socket 3 a click should be heard ensure that it is orientated correctly so that the sockets mate with the corresponding pins on the valve 6 Tighten the strain relief nut 6 and push the electrical supply socket 3 onto the three connector pins on the valve 7 Refit the retaining screw 4 and tighten ensure that there is a slight comp...

Страница 13: ... switching of the electrical supply to the valve Normally open valves close when the electrical supply to the valve is switched on the valve opens automatically when the electrical supply is switched off Normally closed valves open when the electrical supply to the valve is switched on the valve closes automatically when the electrical supply is switched off WARNING Do not touch the surface of the...

Страница 14: ...by the user Carry out the following checks when maintaining the vacuum system Check that the connection to the vacuum system and to the vent gas supply if fitted is secure Check that the electrical connections are secure and that the electrical cable is not damaged WARNING Do not inhale or touch the breakdown products of fluorinated materials which may be present if the vacuum system has been heat...

Страница 15: ... disposal 6 1 Storage Place protective covers over the valve ports and store the IPVA10EK valve in cool dry conditions until required for use When required prepare and install the valve as described in Section 3 6 2 Disposal Dispose of the valve and any components removed from it safely in accordance with all local and national safety and environmental requirements ...

Страница 16: ...C417 60 880 Issue A Page 10 Edwards Limited 2013 All rights reserved Edwards and the Edwards logo are trademarks of Edwards Limited This page has been intentionally left blank ...