Page 9
W603200880_A - Technical data
Technical data
Table 1
General
Measurement range
"Validity"
Accuracy
*
*
Non linearity, hysteresis, repeatability at 25 °C ambient operating temperature without
temperature effects after operation of 2 hours.
0.15% of reading
Temperature effect on zero
0.1 F.S.
0.0050% F.S./ °C
1.0 to 1100 F.S.
0.0025% F.S.
Temperature effect on span
0.01% of reading /°C
Resolution
0.003% F.S.
Gas type dependence
None
Mass
837 to 897 grams
Table 2
Output signal
Output signal analogue (measurement
signal)
Voltage range
-5 to +10.24 V (limited to +10.24 V)
Measurement range
0 to +10 V
Relationship voltage-pressure
Linear
Output impedance
0
(short-circuit proof)
Loaded impedance
> 10 k
Response time
*
*
Increase 10 to 90% F.S.R.
0.25 Torr/mbar (F.S.)
30 ms
0.1 Torr/mbar (F.S.)
130 ms
Table 3
Gauge identification
Resistance R
ident
Voltage
Resistance 13.2 k
referenced to supply
common
5 V
Table 4
Switching functions
Switching functions
SP1, SP2
Setting range
0 to 99% F.S. (0 to 9.9 volts)
Hysteresis
1% F.S.
Relay contact
30 V d.c. /
0.5 A (d.c.)
floating (n.o.)
Closed
p
p
sp
(LED is illuminated)
Open
p
p
sp
(LED is not illuminated)
Switching time
50 ms