Inlet connection
NW40
Outlet connection
NW25
Degree of protection (IEC60529)
IP44
3.4 Electrical data
Pumps are supplied set to 200
‑
230 V.
Table 5
Electrical data
Supply (V)
Phase
Frequency (Hz)
Current (A)
200
‑
230
Single
50 - 60
6.6
100
‑
120
Single
50 - 60
10.6
Table 6
Recommended regional supply protection
Area
Voltage
Rating
UK
230 V
13 A
Europe
230 V
16 A
US
120 V
15 A
Japan
100 V
15 A
3.4.1 Electrical cables
Recommended cordsets and fuses for regional requirements.
Table 7
Recommended cordsets
Description
Rating
Coupler type
Item number
Cord set
assembly, UK
Cable Style = H05VV
‑
F, 3 x 1.5 mm
2
,
300 V, 70 °C, maximum length of
2.5 metres
Straight entry A50505003
Plug Type = BS1363 UK plug
Appliance coupler = IEC60320 style
C19
Right angled
entry
A50505006
Fuse Type = BS1362 13 amp fuse, to
an IEC60320 style
Cord set
assembly, Europe
Cable Style = H05VV
‑
F, 3 x 1.5 mm
2
,
300 V, 70 °C, maximum length of
2.5 metres
Straight entry A50506003
Plug Type = European Schuko VDE
approved, 16 A 250 V rated with
dual earthing contact
Appliance coupler = IEC60320
style C19
Right angled
entry
A50506006
Cord set
assembly,
Cable style = SJT, 3 x 14 AWG, 300 V,
90 °C, VW
‑
1 maximum length of
3 metres
N/A
Page 13
A73101880_F - Technical data
Содержание A73101983
Страница 19: ...Figure 3 Installation drawing dimensions in mm Page 17 A73101880_F Installation ...
Страница 38: ...This page has been intentionally left blank ...