Technical data
A705-40-880 Issue B
2.6
Temperature control system
2.7
Cooling-water data
WARNING
If your pump is part of an ATEX system, it may have an operating case temperature limit. Refer to
the ATEX System instruction manual for details.
Table 6 - Temperature control system data
Water cooling system
Type
Indirect water-to-coolant heat exchanger
Coolant capacity
Refer to
TCV (Thermostatic Control valve)
10 to 90 ºC
Sensor operating temperature range
50 to 175 ºF
Maximum sensor temperature
130 ºC
265 ºF
Thermal snap-switch
Operating temperature range
50 to 100 ºC
120 to 212 ºF
Contact rating
Maximum voltage
240 V
Maximum current (inductive load)
5 A (inductive) 12 V (resistive)
Table 7 - Cooling-water data
Supply temperature range
5 to 35
°C
40 to 95 °F
Maximum supply pressure
10 bar, 1 x 10
6
Pa, 145 psi g
Minimum required pressure differential
across supply and return
2.1 bar, 2.1 x 10
5
Pa, 30 psi g
EDP80
EDP160
EDP250
EDP400
50 Hz
60 Hz
50 Hz
60 Hz
50 Hz
60 Hz
50 Hz
60 Hz
Typical heat removed from pump
by cooling-water
2.5 kW
8540
BTU/h
3 kW
10240
BTU/h
4 kW
13660
BTU/h
4 kW
13660
BTU/h
5 kW
17100
BTU/h
5 kW
17100
BTU/h
5 kW
17100
BTU/h
5 kW
17100
BTU/h
Maximum water consumption
Pump operating temperature
55 °C/131 ºF
150 l/h
40 US
gal/h
150 l/h
40 US
gal/h
270 l/h
72 US
gal/h
270 l/h
72 US
gal/h
270 l/h
72 US
gal/h
270 l/h
72 US
gal/h
330 l/h
84 US
gal/h
330 l/h
84 US
gal/h