Technical data
A373-10-880 Issue R
2
Technical data
2.1
Operating and storage conditions
Note:
To comply with CSA standards, the pump must be installed and used indoors, and within the operating
conditions specified in
below.
2.2
Performance
Note:
Where total pressures are shown in
below, measurements were taken using an untrapped total
pressure capacitance diaphragm gauge on a header, as specified by Pneurop standards.
Table 1 - Operating and storage conditions
Parameter
Reference Data
Ambient temperature range (operation)
13 to 40 °C
Ambient temperature range (storage)
-30 to 70 °C
Normal surface temperature of the pump-body
*
*
At ultimate vacuum, with ambient temperature of 20 °C.
45 to 65 °C
Maximum humidity (operation)
90 % RH
Maximum altitude (operation)
2000 m
Pollution degree
2
Installation category
II
Noise level at 1 metre
57 dB(A) (at 50 Hz)
Vibration severity
†
†
Measured at the inlet port to ISO 2372 (1974)
Class 1C
Table 2 - Performance data
Maximum displacement
E2M28
E2M30
50 Hz electrical supply
32.2 m
3
h
-1
19 cfm
60 Hz electrical supply
38.9 m
3
h
-1
23 cfm
Maximum pumping speed - Pneurop
50 Hz electrical supply
27.5 m
3
h
-1
16.2 cfm
60 Hz electrical supply
33.0 m
3
h
-1
19.4 cfm
Motor rotational speed
50 Hz electrical supply
1440 r min
-1
1440 r min
-1
60 Hz electrical supply
1720 r min
-1
1720 r min
-1
Ultimate vacuum
without gas-ballast (partial pressure)
1 x 10
-4
mbar
1 x 10
-2
Pa
7 x 10
-5
torr
without gas-ballast (total pressure)
1 x 10
-3
mbar
1 x 10
-1
Pa
7 x 10
-4
torr
with full gas-ballast (partial pressure)
2 x 10
-1
mbar
2 x 10
1
Pa
1.5 x 10
-1
torr
Maximum water vapour inlet pressure
30 mbar
3 x 10
3
Pa
22.5 torr