EDTM TINT-CHEK PRO TC3800 Скачать руководство пользователя страница 1

CALIBRATION STANDARDS (PART# TC3811)

Two  light  transmission  test  standards  accompany  each  meter.  The 
standards have VLT values of approximately 27% and 59%.   The actual 
calibrated  value  of  each  standard  is  indicated  on  each  sample.  The 
standards were created using a NIST traceable source.  Use the standards 
to verify that your meter measures the VLT values /- 2% of their 
indicated values. You should regularly verify that your meter is operating 
within specifications.   Before using the standards, ensure they are clean 
and not scratched.  Blow off any dust that may appear on the surface, and 
remove any finger prints with a micro-fiber cloth.   If the meter measures 
outside the +/-2% range, please inspect the cleanliness of the reflector area 
of the REFLECTOR unit, and use compressed air to blow off the back side 
of  the  INSTRUMENT  unit  as  well.    If  problems  still  exist,  contact  the 
manufacturer for a possible solution.

When testing with the calibration 
s t a n d a r d s ,   c o n d u c t   y o u r  
measurement  in  the  same 
manner  as  testing  an  actual 
window.  When  not  in  use,  store 
the  calibration  standards  in  the 
re-sealable  bag  they  were 
packaged  in,  and  keep  them 
inside the carrying case for best 
protection.    If  the  calibration 
standards  become  scratched  or 

damaged, replacement sets can be ordered from the manufacturer.   The 
part number for the replacement standard set is TC3811.  

WARRANTY

The manufacturer warrants all models of the TC3800 to be free from defects 
in material and workmanship under normal use and service as specified 
within the operator's manual.  The manufacturer shall repair or replace the 
unit within twelve (12) months from the original date of shipment after the 
unit is returned to the manufacturer's factory, prepaid by the user, and the 
unit is disclosed to the manufacturer's satisfaction, to be thus defective.   
This warranty shall not apply to any unit that has been repaired or altered 
other  than  by  the  manufacturer.   The  aforementioned  provisions  do  not 
extend the original warranty period of the unit which has been repaired or 
replaced by the manufacturer.  Batteries are not covered by warranty.

The manufacturer assumes no liability for the consequential damages of 
any kind through the use or misuse of the TC3800 product by the purchaser 
or others.   No other obligations or liabilities are expressed or implied.   All 
damage or liability claims will be limited to an amount equal to the sale price 
of the TC3800, as established by the manufacturer.

TINT-CHEK PRO

USER GUIDE

Model# TC3800

S

Wavelength............................ 550   Nanometers
Resolution.............................. 0.1%
Accuracy................................ +/- 2 Percentage Points
Repeatability ......................... 1 Percentage Point
Measurement Range............. 0 to 100% Light Transmission
Operating Temperature......... 0 F to 122 F (-17 C to +50 C)
Storage Temperature............ -4 F to 158 F (-20 C to +70 C)
Light source.......................... LED
Powered by ......................... 9V Alkaline Battery (Supplied)
Sample Thickness................ 0 to 0.25 inches  (0 to 6.5 mm)
Minimum Sample Size.......... 1 inch x 1 inch (25 x 25mm)
Instrument Dimensions......... 5.1 x 2.6 x 2.0 inches  (130 x 65 x 52mm)
Instrument Weight................. 0.78 pound (0.35kg)

PECIFICATIONS

BATTERY REPLACEMENT

This instrument is powered by a single 9-volt alkaline battery (included).   
When the battery voltage is getting low, the battery indicator will illuminate.  
The instrument is still operable at this point, however it is recommended 
that the battery be replaced soon.  Alkaline batteries are recommended for 
this product. 

The TINT-CHEK PRO Window Tint Meter tests the Light Transmission % 
(VLT) of automotive windows, as well as flat glass applications.   The two-
piece design allows for testing of the windshield, as well as rear and side 
windows.  The two piece instrument consists of a REFLECTOR unit and an 
INSTRUMENT unit that self-calibrates at start-up.  The INSTRUMENT unit 
contains all of the electronics and is battery powered.   The REFLECTOR 
unit does not contain any electronics and simply houses the reflector and 
magnets to help in aligning the two enclosures together for measurements.

INSTRUMENT

UNIT

REFLECTOR

UNIT

745 Capital Commons Drive

Toledo, Ohio  43615    USA

PHONE: (419) 861-1030

FAX: (419) 861-1031

WWW.EDTM.COM

Email:  [email protected]

TINT-CHEK.COM

EDTM.COM

MADE IN THE USA

Отзывы: