![EDTM TINT-CHEK PRO TC3800 Скачать руководство пользователя страница 1](http://html1.mh-extra.com/html/edtm/tint-chek-pro-tc3800/tint-chek-pro-tc3800_user-manual_2362826001.webp)
CALIBRATION STANDARDS (PART# TC3811)
Two light transmission test standards accompany each meter. The
standards have VLT values of approximately 27% and 59%. The actual
calibrated value of each standard is indicated on each sample. The
standards were created using a NIST traceable source. Use the standards
to verify that your meter measures the VLT values /- 2% of their
indicated values. You should regularly verify that your meter is operating
within specifications. Before using the standards, ensure they are clean
and not scratched. Blow off any dust that may appear on the surface, and
remove any finger prints with a micro-fiber cloth. If the meter measures
outside the +/-2% range, please inspect the cleanliness of the reflector area
of the REFLECTOR unit, and use compressed air to blow off the back side
of the INSTRUMENT unit as well. If problems still exist, contact the
manufacturer for a possible solution.
When testing with the calibration
s t a n d a r d s , c o n d u c t y o u r
measurement in the same
manner as testing an actual
window. When not in use, store
the calibration standards in the
re-sealable bag they were
packaged in, and keep them
inside the carrying case for best
protection. If the calibration
standards become scratched or
damaged, replacement sets can be ordered from the manufacturer. The
part number for the replacement standard set is TC3811.
WARRANTY
The manufacturer warrants all models of the TC3800 to be free from defects
in material and workmanship under normal use and service as specified
within the operator's manual. The manufacturer shall repair or replace the
unit within twelve (12) months from the original date of shipment after the
unit is returned to the manufacturer's factory, prepaid by the user, and the
unit is disclosed to the manufacturer's satisfaction, to be thus defective.
This warranty shall not apply to any unit that has been repaired or altered
other than by the manufacturer. The aforementioned provisions do not
extend the original warranty period of the unit which has been repaired or
replaced by the manufacturer. Batteries are not covered by warranty.
The manufacturer assumes no liability for the consequential damages of
any kind through the use or misuse of the TC3800 product by the purchaser
or others. No other obligations or liabilities are expressed or implied. All
damage or liability claims will be limited to an amount equal to the sale price
of the TC3800, as established by the manufacturer.
TINT-CHEK PRO
USER GUIDE
Model# TC3800
S
Wavelength............................ 550 Nanometers
Resolution.............................. 0.1%
Accuracy................................ +/- 2 Percentage Points
Repeatability ......................... 1 Percentage Point
Measurement Range............. 0 to 100% Light Transmission
Operating Temperature......... 0 F to 122 F (-17 C to +50 C)
Storage Temperature............ -4 F to 158 F (-20 C to +70 C)
Light source.......................... LED
Powered by ......................... 9V Alkaline Battery (Supplied)
Sample Thickness................ 0 to 0.25 inches (0 to 6.5 mm)
Minimum Sample Size.......... 1 inch x 1 inch (25 x 25mm)
Instrument Dimensions......... 5.1 x 2.6 x 2.0 inches (130 x 65 x 52mm)
Instrument Weight................. 0.78 pound (0.35kg)
PECIFICATIONS
BATTERY REPLACEMENT
This instrument is powered by a single 9-volt alkaline battery (included).
When the battery voltage is getting low, the battery indicator will illuminate.
The instrument is still operable at this point, however it is recommended
that the battery be replaced soon. Alkaline batteries are recommended for
this product.
The TINT-CHEK PRO Window Tint Meter tests the Light Transmission %
(VLT) of automotive windows, as well as flat glass applications. The two-
piece design allows for testing of the windshield, as well as rear and side
windows. The two piece instrument consists of a REFLECTOR unit and an
INSTRUMENT unit that self-calibrates at start-up. The INSTRUMENT unit
contains all of the electronics and is battery powered. The REFLECTOR
unit does not contain any electronics and simply houses the reflector and
magnets to help in aligning the two enclosures together for measurements.
INSTRUMENT
UNIT
REFLECTOR
UNIT
745 Capital Commons Drive
Toledo, Ohio 43615 USA
PHONE: (419) 861-1030
FAX: (419) 861-1031
WWW.EDTM.COM
Email: [email protected]
TINT-CHEK.COM
EDTM.COM
MADE IN THE USA