The SP2065 is designed to operate in either power mode or transmission mode. While
operating in power mode, the SP2065 is identical to our SP1065 meter. However, when the
SP2065 is switched into transmission mode, the meter self-calibrates to read 100%
transmission for the current solar conditions presented to the meter. Thereafter all readings are
referenced against the solar conditions that were present when the self-calibration occurred.
Therefore if you place a piece of glass that passes 40% of the solar energy between the meter
and the light source, the display will read P 40 (i.e. 40 % solar transmission).
The SP2065 is helpful in demonstrating the performance of fenestration products. Even more
importantly, the SP2065 can be used to identify types of Low E coatings in the field. By knowing
the general performance of specific coatings, the user can differentiate types of coatings
MEASURING INSTALLED WINDOWS
The best results are obtained when the meter is operated on windows that are directly
facing the sun. The position of the meter is EXTREMELY important in obtaining an
accurate reading. The meter should be placed on the window sill on top of it's stand or
other stationary item that will hold the meter in the same position. To take a transmission
reading, the meter should be self-calibrated with the window open (Step 3 & 4 above).
The sensing area of the meter should NOT be looking through any glass or film at this
time. In step #5 of the instructions, you should close the window without moving the
position of the meter. Once the window is closed, the meter will measure the solar
energy that is transmitting through the window. Several readings should be taken to
guarantee the most accurate information is obtained.
Please note that outdoor
conditions are ever-changing. Clouds and environmental conditions can change
the incident solar energy quickly.
SAMPLE BEING TESTED
STEPS 2 - 4
STEP 5 & 6
% TOGGLE
BUTTON
TABLETOP DEMONSTRATIONS/QC WORKSTATIONS
To set up a
TABLETOP DEMONSTRATION
, or to perform readings in a bench-type QC
application, place a light source facing the SP2065. The light source should be
stable and locked in position so it cannot move during the measurements. Position
the meter flat on the table with the sensor facing the light source directly. If
necessary, place the SP2065 on top of a stand (#SP2075) to position the meter in the
center of the light beam. To obtain the most accurate results, DO NOT move the light
source OR the meter during the readings. To take solar transmission measurements,
follow the steps below and the illustrations on the next page:
1) Turn on the SP2065 by pushing the power button
2) Place the meter on the table or platform with the light sensor (top end of
meter) facing the light source directly. Turn on the light source and confirm
that a power reading has registered on the display.
3) Self-calibrate the meter by pressing the BLUE % switch next to the power button.
4) The display should now read P100 (=100% transmission). Your meter is now
prepared to take a transmission measurement with the current light
conditions. (If the surrounding light conditions change, repeat STEP 3).
5) Place the window or film sample between the light source and the SP2065.
6) The resulting solar transmission percentage for that material will be displayed.
7) Remove the sample and confirm the SP2065 returns to P100. If the meter
does not display P100 with the sample removed, disregard the reading and
begin the test process again.
INFRARED HEAT LAMP
EDTM MODEL# HL1040
LIGHT SENSOR
sp2065_120829.cdr
FIGURE 1
MINIMUM 18”
ACCESSORIES
For more information and a complete list visit EDTM.com.
#HL1040
Heat Lamp
#SP2075
Stand
#HS2056
Demonstration Base
CHANGING THE UNIT OF MEASURE
2
2
The SP2065 accommodates measurements in units of both Watts/Meter and BTU/hour*foot .
At start up, the meter will indicate the current setting on the LCD display (See Figure 1), as well
as by indicating the unit of measure by the LED indicators. The meter will retain the previously
established unit of measure until it is modified by the user. Therefore you do not have to
modify it each time the meter is powered on. Follow the instructions below to change the unit
of measure.
1) Power the unit on and release the power button.
2) Press
AND HOLD
the power button as though you are turning it off.
3) The display will go blank as though the instrument is turning off.
4) Continue to hold down the power button for 3 seconds.
5) The display will briefly show the current setting, and then switch to the new setting.
6) The LED indicator will also switch to the new setting to confirm your choice.
7) When the screen shuts off again release the power button.
8) The next time you turn the meter on, it will function in the new unit of measure.
2
Indicates BTU/hr*ft
2
Indicates W/m