The  SP2065  is  designed  to  operate  in  either  power  mode  or  transmission  mode.  While 
operating in power mode, the SP2065 is identical to our SP1065 meter.   However, when the 
SP2065  is  switched  into  transmission  mode,  the  meter  self-calibrates  to  read  100% 
transmission for the current solar conditions presented to the meter.  Thereafter all readings are 
referenced against the solar conditions that were present when the self-calibration occurred.  
Therefore if you place a piece of glass that passes 40% of the solar energy between the meter 
and the light source, the display will read P 40 (i.e. 40 % solar transmission).

The SP2065 is helpful in demonstrating the performance of fenestration products.  Even more 
importantly, the SP2065 can be used to identify types of Low E coatings in the field.  By knowing 
the general performance of specific coatings, the user can differentiate types of coatings

MEASURING INSTALLED WINDOWS

The best results are obtained when the meter is operated on windows that are directly 
facing  the  sun.    The  position  of  the  meter  is  EXTREMELY  important  in  obtaining  an 
accurate reading.  The meter should be placed on the window sill on top of it's stand or 
other stationary item that will hold the meter in the same position.  To take a transmission 
reading, the meter should be self-calibrated with the window open (Step 3 & 4 above).  
The sensing area of the meter should NOT be looking through any glass or film at this 
time.   In step #5 of the instructions, you should close the window without moving the 
position of the meter.   Once the window is closed, the meter will measure the solar 
energy that is transmitting through the window.   Several readings should be taken to 
guarantee  the  most  accurate  information  is  obtained.   

Please  note  that  outdoor 

conditions are ever-changing.  Clouds and environmental conditions can change 
the incident solar energy quickly.

SAMPLE BEING TESTED

STEPS 2 - 4

STEP 5 & 6

% TOGGLE 

BUTTON

TABLETOP DEMONSTRATIONS/QC WORKSTATIONS

To set up a 

TABLETOP DEMONSTRATION

, or to perform readings in a bench-type QC 

application, place a light source facing the SP2065.   The light source should be 
stable and locked in position so it cannot move during the measurements.  Position 
the  meter  flat  on  the  table  with  the  sensor  facing  the  light  source  directly.    If 
necessary, place the SP2065 on top of a stand (#SP2075) to position the meter in the 
center of the light beam.  To obtain the most accurate results, DO NOT move the light 
source OR the meter during the readings.  To take solar transmission measurements, 
follow the steps below and the illustrations on the next page:

1) Turn on the SP2065 by pushing the power button
2) Place the meter on the table or platform with the light sensor (top end of
    meter) facing the light source directly.  Turn on the light source and confirm
    that a power reading has registered on the display.
3) Self-calibrate the meter by pressing the BLUE % switch next to the power button.
4) The display should now read P100 (=100% transmission). Your meter is now
    prepared to take a transmission measurement with the current light
    conditions.  (If the surrounding light conditions change, repeat STEP 3).
5) Place the window or film sample between the light source and the SP2065.
6) The resulting solar transmission percentage for that material will be displayed.
7) Remove the sample and confirm the SP2065 returns to P100.  If the meter
    does not display P100 with the sample removed, disregard the reading and
    begin the test process again.

INFRARED HEAT LAMP

   EDTM MODEL# HL1040

LIGHT SENSOR

sp2065_120829.cdr

FIGURE 1

MINIMUM 18”

ACCESSORIES

For more information and a complete list visit EDTM.com.   

#HL1040

Heat Lamp

#SP2075

Stand

#HS2056

Demonstration Base

CHANGING THE UNIT OF MEASURE

2

2

The SP2065 accommodates measurements in units of both Watts/Meter  and BTU/hour*foot .  
At start up, the meter will indicate the current setting on the LCD display (See Figure 1), as well 
as by indicating the unit of measure by the LED indicators.  The meter will retain the previously 
established unit of measure until it is modified by the user.   Therefore you do not have to 
modify it each time the meter is powered on.  Follow the instructions below to change the unit 
of measure.

1) Power the unit on and release the power button.
2) Press 

AND HOLD

 the power button as though you are turning it off.

3) The display will go blank as though the instrument is turning off. 
4) Continue to hold down the power button for 3 seconds.
5) The display will briefly show the current setting, and then switch to the new setting.
6) The LED indicator will also switch to the new setting to confirm your choice.
7) When the screen shuts off again release the power button.
8) The next time you turn the meter on, it will function in the new unit of measure.

2

Indicates BTU/hr*ft

2

Indicates W/m