Section 2 -

 

Calibration and Measurement

For accurate results, stir all buffers and samples. Always rinse electrodes and blot dry 
before transferring from one solution to another to prevent contamination. Ensure that 
any electrode filling holes are left uncovered during use.

Automatic  Calibration & Temperature Compensation
1.

 Connect the pH electrode(s) and temperature probe and  switch on using the I/O Key.

2.

 Select pH mode and clear any calibration data by pressing the Clear key for 5 seconds. 

3.

 Place the pH electrode(s) and the temperature probe in the pH 7 buffer and wait for a 

stable reading. Press the BUF 1 key. The first calibration point will now be entered 
automatically at the correct value for the temperature of measurement. The Cal flag will 
be displayed.

4.

 Rinse, blot and place the pH electrode(s)  and temperature probe in the second buffer 

(4 or 10). Wait for a stable reading and press BUF 2. The second calibration point will  
now be entered automatically at the correct value for the temperature of measurement. 

5. 

Rinse, blot and place pH electrode(s) and temperature probe in the sample. Wait for a 

stable reading and record pH value.

6.

 Temperature or mV measurements may be made at any time by pressing the mode 

key.

7.

 To adjust the E° value of the electrode, place in buffer 4,7 or 10 and press the BUF 1 

key.

Note:

 The performance of the electrode may be monitored by pressing the SLOPE key. 

The slope is displayed as a % of Nernstian. The value should lie between 85 and 110%.

Manual Calibration

If buffers other than 4,7 and 10pH are to be used, then the calibration must be performed 
manually. That is the correct pH values must be entered using the ñò keys
Follow points 

1

 and 

2

 for Auto calibration above : 

3.

 Place the electrode(s) and temperature probe in the first buffer and wait for a stable 

reading. Check the temperature of the buffer. Look up the value of the buffer for this 
temperature

4. 

Use the       keys to adjust the displayed reading to the value displayed in 

3.

 (the CAL flag 

will flash). Press BUF 1 (the CAL flag will stop flashing)

5.

 Place the electrode(s) and temperature probe in the second buffer and wait for a stable 

reading. Check the temperature of the buffer. Look up the value of the buffer for this 
temperature

6.

 Use the       keys to adjust the displayed reading to the value displayed in 

5.

 (the CAL flag 

will flash). Press BUF 2 (the CAL flag will stop flashing)

7. 

Rinse, blot and place pH electrode(s) and temperature probe in the sample. Wait for a 

stable reading and record pH value.

8.

 Temperature or mV measurements may be made at any time by pressing the mode key.

9.

 To adjust the E° value of the electrode, place in buffer 4,7 or 10 and press the BUF 1 key.

Содержание FE257

Страница 1: ...et Up Section 2 pH Calibration Measurement Section 3 mV Calibration Measurement Section 4 Use of the Recorder Output Section 5 Troubleshooting Appendix 1 pH Electrodes Buffers Section 6 Specification...

Страница 2: ...t Note The carrier will not honour any claims unless all shipping material is saved for their examination After examining and removing contents save packing material in the event that re shipment be n...

Страница 3: ...ue in the range 0 0 to 19 0 C and is adjustable i e responds to the use of the keys Adjust to 20 0 C 6 The meter is ready for calibration A Note on Electrodes pH Ion Selective and Redox electrodes may...

Страница 4: ...ust the E value of the electrode place in buffer 4 7 or 10 and press the BUF 1 key Note The performance of the electrode may be monitored by pressing the SLOPE key The slope is displayed as a of Nerns...

Страница 5: ...i e Section 3 mV Calibration and Measurement The mV mode has two ranges The meter will automatically select the most appropriate range as follows 0 to 400mV with resolution 0 1mV 400mV to 2000mV with...

Страница 6: ...n left hand digit Over range condition Drifting readings Erratic drifting readings or display reads on left hand side when shorting plug is attached Section 5 Troubleshooting Probable Cause Battery is...

Страница 7: ...electrode storage solution is made up by dissolving 1g KCl in 100mL of pH 7 buffer Buffer Capsules Buffer capsules are made up as follows For each buffer empty the powder into a suitable container Usi...

Страница 8: ...e information on our products please visit our website www edt co uk EDT directION Unit 5 Waldershare Park Dover CT15 5DQ Telephone 01304 829960 Email sales edt co uk Company No 04135318 VAT No GB 765...

Отзывы: