2-1.2  Operation of Trendsetter™ with LVDT Electronic Probes
Using the power switch located at the rear of the column, turn the unit on. Depending upon your
application, select the input channel as required:  A, B, or AB. Next, check the Inch, Metric, indica-
tor light to see if the gage is set to the required measurement system. Depress the contacts on the
probe to see if the column moves in the correct direction for the movement of the contacts; i.e., for
an O.D. measurement, a large part will cause the probe tip to move in. Any depression on the probe
tip should move the bargraph display in an upward direction. If either of these functions is incorrect,
remove the module as follows:  switch the column off, unscrew the two module retraining screws,
and gently pull the module out. Do not touch the matrix I/O pin headers; the function of these units
can be found later in this write-up.

Check that the Inch and Metric Pin Headers as well as the polarity switches are in their correct
position. Replace the module by lining up the top and bottom edges of the board with it’s card
guides and sliding back until it locates in its electrical socket. Screw in the two module retaining
screws to secure the module and switch power on. The full scale range of the readout should now
be set.

To select a range, if it has not already been decided, it is best to select one that is two to three
times greater than the tolerance being checked. This will allow for good approach and oversize
range. If parts are being matched or segregated into classes and are all in tolerance, much more of
the full scale range can be used.

Set the range knob to the desired value. The corresponding digital scale values will be shown
alongside the bargraph scale. The setup is now ready for mastering.

2-1.3  Mastering
There are two basic approaches to mastering:

1.)  Min. and Max. masters
2.)  Mean or zero master

Setting to a Mean Master
Place the mean master in a fixture and turn the zero knob to bring the scale to zero or to the actual
calibrated size of the master.

Setting to Min. & Max. Masters
Place the max. master in the gage and, with the zero knob, adjust the column to the upper limit
reading. Remove the max. master and replace with the min. master. The reading should rise to the
lower limit. If, due to a poor initial calibration (or a non 1:1 ratio in the lever arms of the gage), the
lower limit is not achieved, the magnification should be altered.

2-1.4  Front Panel Magnification Trim
Place one master in the gage. Set one limit at the desired point with the zero knob. Apply the other
master and, with a small screwdriver, adjust the recessed “mag” screw until the column moves to
the desired point. Reapply the other master and, if necessary, readjust until both masters move the
column the required amount. This front adjust screw will allow for a trim of approximately 20% and
will change all inputs; A, B, and AB.

In some cases, where transducers other than Edmunds’ gage probes are being used, or where
levers change the 1:1 ratio of input to output, the rear adjustments on the circuit board may need to
be used.

5

Содержание TRENDSETTER

Страница 1: ...INSTRUCTION MANUAL Edmunds GAGES EDMUNDS GAGES FARMINGTON INDUSTRIAL PARK FARMINGTON CT 06032 USA 860 677 2813 FAX 860 677 4243 Edmunds TRENDSETTER...

Страница 2: ...ease you ve come to expect from all Edmunds products The overwhelming success of the Trendsetter reflects a company wide commitment to continual technological improvement technical integration and cus...

Страница 3: ...fying Module 16 3 4 E8011 Limit Light Module 5 Point 18 3 5 E8023 Air Electronic Module 20 3 5 1 Air Electronic Maintenance 22 3 5 2 Block Diagrams E8023 24 3 6 E8024 T I R Module with Limit Lights 25...

Страница 4: ...manual To form a firm three point support for the column unscrew the front screw at the bottom of the long foot and remount the foot on the front screw The foot should be perpendicular to the column...

Страница 5: ...llow for good approach and oversize range If parts are being matched or segregated into classes and are all in tolerance much more of the full scale range can be used Set the range knob to the desired...

Страница 6: ...erns differential measure ments when both probes are measuring a diameter for example Also lack of balance is one of the chief causes of poor system repeatability Balance is easily achieved and observ...

Страница 7: ...escribed in the preceding section Repeat the procedure several times until no movement is observed on the Trendsetter display Restore the system to its original settings and recheck cali bration with...

Страница 8: ...on ranges low medium and high It can be run in inch or Metric units with the high and low readings either at the top or bottom of the scale The pin headers to control these functions are located on th...

Страница 9: ...ch to required posi tion using a small screwdriver inserted thru the front panel 2 2 3 Mastering 1 Check the two master sizes and determine the set points on the bargraph scale With an air plug the sm...

Страница 10: ...em tooling which has a very shallow nozzle drop will show very high magnification when used on a Trendsetter This might necessitate a change to a lower air amplification on the 3 range pin header Diff...

Страница 11: ...n is 35 watts maximum 15 watts is more typical and will drive a full compli ment of plug in modules A 10 inch 101 point bargraph display is the pri mary means of readout The 101 points give a readable...

Страница 12: ...output connector and D C common to P11 and adjust gain pot for full scale 6 Repeat step 5 with 2 500 volt and readjust if necessary 7 Repeat step 4 6 until full scale and center zero have been set Rep...

Страница 13: ...or may not be used as required See matrix switching and examples Rear Panel 25 Pin I O Connector Assignments Pin Number Input Connector Output Connector 1 Analog Output 1 2 Analog Output 2 3 Analog Ou...

Страница 14: ...ntly used position of the mode switch When in this position the limit lights over good under follow the instantaneous value of the lower bay module This position disables the remote interface 4 Auto R...

Страница 15: ...ate these special modes turn the function switch to Auto Run position Determine the desired function and using the truth tables apply a 5 24 VDC command to the I O pins listed These signals must be po...

Страница 16: ...supply s positive output is connected to Pin 9 of the module The power supply ground is referenced to the load Controls Adjustments The Low and High set adjustments on the front of the E8007 module re...

Страница 17: ...g the Zero knob on the TrendsetterTM lower bay module set the bargraph display to the lowest position on scale 7 Using the Zero knob on the TrendsetterTM lower bay module set the bargraph display to t...

Страница 18: ...o ensure a successful operation the low setpoint must be less than the low approach setpoint the low approach setpoint must be less than the upper approach setpoint the upper approach setpoint must be...

Страница 19: ...tive with respect to user common P24 0 Off State 1 On State Truth Table Rear Panel I O P15 Write Disable P16 Reset Function Off 0 Off 0 Follow Default Off 0 On 1 Latch On 1 On 1 Hold On 1 Off 0 Reset...

Страница 20: ...sition pin header allowing access to the I O busses and a 5 position range encoding switch to signal the readout display Two 3 8 bulkhead fittings provide a convenient means for attaching the air supp...

Страница 21: ...tooling As a rule of thumb attempt to calibrate the air tooling in the low position If you are unable to set the magnification or if the dynamic response of the air signal is too slow use the medium p...

Страница 22: ...eck the mastering points before gaging Bias Restriction Remove bias bleed screw and metallic filter Insert 1 4 28 screw into dropping resistor and pull out restrictor assembly Remove bottom metallic f...

Страница 23: ...an the existing slot and do not over bend With the valve assembled and all O rings in position reinsert the valve taking care to go slowly so as not to nick the top O ring as it passes the set screw h...

Страница 24: ...24...

Страница 25: ...d to display either a or peak signal This feature is equally useful as TIR and is typically used as follows In matching an I D and an O D out of round is always present to some extent The E8024 module...

Страница 26: ...Install one of the masters in the gage fixture so that the bargraph is on scale Rotate the Mode Switch to the TIR position The bargraph scale Zero will automatically move to the bottom posi tion This...

Страница 27: ...ory Rotate the part 360 degrees The dis played readings will be the average of the readings obtained during part rotation Remote Interface Two user control lines present on the I O connectors located...

Страница 28: ...e rear panel of the Trendsetter column The outputs are open collector NPN OPTO transistors whose emitters are referenced to user common Pin 24 The transistors may be pulled to a maximum of 24 VDC OUTP...

Страница 29: ...d in match gaging here the I D and O D signals from each of two columns are fed into the sum and difference module and the result displayed as clearance The inputs are divided into 2 groups of three A...

Страница 30: ...ut groups will be summed together A B or operated in the difference mode A B When used in conjunction with the polarity pin header any of the four combinations listed below may be established A B A B...

Страница 31: ...e six inputs of the module The mag pots have an attenuation of 25 1 this allows for mixing of signals with wide differences in their magnifications or formats Pin Header Function 1 A1 Group Input 2 A2...

Страница 32: ...32...

Страница 33: ...inch 04mm 020 inch 4mm 005 inch 10mm 050 inch 1 0mm 010 inch 20mm 100 inch 2 0mm 020 inch 40mm 200 inch 4 0mm Zero Controls A B The two ten turn zero controls provide independent zeroing of the A and...

Страница 34: ...10 of I O lines Combinations can be assigned by connecting the cap to straddle any two of three pins with the signal being present on the center pin see pin cap example Unused caps should be stored o...

Страница 35: ...35...

Страница 36: ...nt allows for 20 change in the overall system magnification The control has a span of 20 turns for precise mag setting Circuit Board Level Selections Polarity Reversal Jumpers Two pin cap jumpers are...

Страница 37: ...display to zero on bargraph 5 Place a known calibrated mechanical load onto the transducer bridge and observe A Did the bargraph move in the proper direction If not reverse the norm rev pin caps B Di...

Страница 38: ...38...

Страница 39: ...r measurement ranges 001 through 0001 Full Scale Inch Metric operation Zero control Manual hold Reset functions Three operating modes CONTROLS Mode Switch Selects one of the three operating modes BYPA...

Страница 40: ...ector Time Select Using the internal jumper selects the time interval the peak detector will hold the measured results before automatically resetting Selectable from 1 2 3 or 5 second intervals or use...

Страница 41: ...e be transmitted or bussed on any of the 10 matrix buss lines dependent on the switch closed For example closing switch 1 outputs the display signal on matrix line 1 The display output volt age is 2 5...

Страница 42: ...ule the bargraph will rise with increasing air pressure at the tooling noz zles in the setting Using electronic modules the bargraph will rise when the tip of the probe is pressed toward the body of t...

Страница 43: ...43...

Страница 44: ...44...

Страница 45: ...the I D of the piece part increases place the polarity pin header so that the bargraph also moves is a positive upscale direction The I D signal of this module will be placed on buss 2 by placing a pr...

Страница 46: ...will be programmed for A B operation receive 2 inputs and display the result as squareness The first step is to program column 1 matrix pin headers Program matrix pin headers AO and BO onto the system...

Страница 47: ...47...

Страница 48: ...48...

Страница 49: ...ne controls external lamps etc The relay interface cabinet receives the sta tus of these outputs decodes the status dis plays the status with its own three limit light dis play and then drivers operat...

Страница 50: ...the corresponding Under and Over relays will be activated and the Good relay will be deactivated Master Accept or master Reject may be implemented by this feature Connections for access to the semiaut...

Страница 51: ...O O ther Edmunds Gages Products and Services Edmunds GAGES Farmington Industrial Park Farmington CT 06032 USA TEL 860 677 2813 FAX 860 677 4243...

Отзывы: