Service Manual

 

 

EMS & FMS Stimulators

 

VG-119-D / 21-Jan-2017 

 

Page 7 of 16 

Power Consumption

 (at max luminance) 

 

+12 VDC 

1.5 A 

 

+24 VDC 

0.5A 

Performance and Operating Specifications 

 

Internal high-resolution, color, LCOS stimulus display 

 

1280 x 1024 pixels at 60 Hz, 75 Hz, or 85 Hz 

 

40 microsecond response time 

 

24-bit RGB color (8 bits per channel) 

 

LED life expectancy of 1,000 hours 

 

Red: 627 nm typical, half-width 20 nm at 350 mA and 25 °C 

 

Green: 530 nm typical, half-width 30nm at 350 mA and 25 °C 

 

Blue: 470 nm typical, half-width 20nm at 350 mA and 25 °C 

 

Approximately 50 degree diameter field of stimulation. 

 

Greater than 300 cd/m2 maximum luminance. 

Infrared Video Eye Monitoring (FMS IV and EMS IV) 

 

Both use IR LEDs: 850 nm dominant wavelength; half-width 45 nm at ½ intensity 

 

Eye camera provides magnified IR image of eye for accurate alignment and monitoring 
during testing. 

 

Infrared fundus camera provides IR fundus image for accurate fixation monitoring during 
testing. 

 

Software provides for capturing and saving single frames of the fundus image.  Recorded 
waveform traces later can be easily superimposed upon saved fundus image images. 

Patient Refraction System 

 

Single-knob correction of spherical refractive errors with automatic compensation for 
magnification/minification stimulus on the subject's retina 

o

 

Maximum myopic correction:   

-20 D SPH 

o

 

Maximum hyperopic correction:    +8 D SPH 

 

Stimulator is mounted on an articulated arm for easy positioning (arm included). 

 

Spot Calibrator mounts on refractor lens for convenient re-calibration using 

VERIS

 

software The Spot Calibrator is included in every Veris™ platform 

Fixation monitoring with IR fundus camera (with FMS stimulators only

 

Stimulator incorporates an IR fundus camera and an IR illumination system of the fundus 

 

Fundus illumination uses IR LEDs at 850 nm dominant wavelength with 45 nm half-
width. 

 

The camera lens has an adjustable aperture. 

 

A Grid Calibrator is included with every FMS installation. It is a model eye that allows 
calibration of the stimulus grid shown overlayed over the video fundus image. 

 

 

Содержание VERIS EMS IV

Страница 1: ...ulators VG 119 D 21 Jan 2017 Page 1 of 16 VERIS FMS IV EMS IV Stimulators Installation Service Manual Release D Electro Diagnostic Imaging Inc http www veris edi com Tel 650 631 0120 Fax 650 631 0122...

Страница 2: ...Information 3 1 4 Warnings and Precautions 3 2 Labels 5 2 1 Component Labels 5 2 2 Shipping Label 5 3 User Profile 6 4 Product Lifetime 6 5 Components of the FMS IV EMS IV Stimulators 6 6 Theory of O...

Страница 3: ...Medical Electrical Equipment Part 2 General requirements for basic safety and essential performance collateral standard Electromagnetic VERIS has been tested for radiative emissions and shown to meet...

Страница 4: ...he United States Canada and Europe If the VERIS System is used with customer provided components testing for safety is the responsibility of the customer Do not sit on the cart or table CAUTION Federa...

Страница 5: ...as CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnost...

Страница 6: ...nob spherical refractive correction with a range of 20 to 8 diopters 7 Specifications Physical Stimulator unit Approx 18 21 x 7 x 4 25 46 54 cm x 19cm x 6 7 cm powder coated upon aluminum housing Note...

Страница 7: ...tion monitoring during testing Software provides for capturing and saving single frames of the fundus image Recorded waveform traces later can be easily superimposed upon saved fundus image images Pat...

Страница 8: ...ox against its packing slip when it is opened to account for all items Mount the arm on the right front edge of the desktop Mount the Control Box to the left of the Arm Mount Mount the tilting mechani...

Страница 9: ...box labeled FMS Power Out The GearMo serial adapter connects directly to the control box serial port and to the serial port of the Grass 15LT amplifier If replacing a previously installed Keyspan Adap...

Страница 10: ...m 5 Start VERIS open a protocol that uses the stimulator and select a plot 6 From Calibration on the menu bar select Calibrator Calibration and make sure Spot Calibrator is Checked 7 Cover the opening...

Страница 11: ...dow to open the camera controls 7 Adjust the image brightness by means of and contrast controls and by rotating the Grid Calibrator as necessary Fluctuations of image brightness are normal If you can...

Страница 12: ...e lens Apply lens cover when the instrument is not in use If necessary clean lens with eyeglass cleaning materials and soft cotton or micropore cloth 3 Calibration Perform Grid Calibration and Auto Ca...

Страница 13: ...crodisplay control box Keyspan adapter USB port of computer Switch the microdisplay control box OFF wait then and ON again Then run VERIS and enter a recording window It may be necessary to try twice...

Страница 14: ...le 76 6FT Cables to Go 1 Mini display DVI adapter Adapter 03 Apple Computers 1 Stimulator Arm Light IOP7500 500 124 EDI 1 Spot Calibrator CAL Spot 01 International Light EDI 1 Grid Calibrator CAL GRID...

Страница 15: ...15 of 16 14 Cables Schematics Diagrams Calibrator DB 25 Connector Cable 76 02 and 76 6FT Standard Male Male VGA Legacy Cables Cable 76 01 Camera Interface Cable Serial Assembly Keyspan Adapter and Cab...

Страница 16: ...VERIS Service Manual EMS FMS Stimulators VG 119 D 21 Jan 2017 Page 16 of 16 15 Software All stimulator functions are controlled by the VERIS application software...

Отзывы: