Flow Adapter 

(P/N FA-104) 

2.0 LPM

Inhalable and Respirable

Note: Before Operating Check Battery Status. Charging time is 16 hours. Operating time  

is 7 hours for Respirable and 8hrs for Inhalable sampling without a gravimetric filter. 

STEP 1

  

Check Battery Level 

Ensure battery is fully charged:

 Turn monitor on. Press “Enter”. Select “Special Functions” 

 “System Options” “Extended Options” 

 “Battery Status.” The battery status reads greater 

than 6.5 Volts when fully charged.

Up / Down

Select/Enter

 

On/Off 

Battery Level

For Inhalable Air Sampling

Ensure switch position is in the upward 

setting, labeled “Inhalable.” If using 

gravimetric filter insert into cassette. 

Flow can be calibrated by inserting the 

Flow adapter (P/N FA-104) into the IOM 

inlet. Select “Run” from the main menu 

to activate air sampling pump. In the 

“Inhalable Flow Adjust” position, flow 

can be adjusted using the trim stick.

Switch Position

Up for Inhalable

Down for Respirable

Inhalable

 Flow Adjust 

Respirable

 Flow Adjust 

Trim 

Stick

STEP 2 

 

Set Flow Rate 

B. For Respirable Air Sampling 

Ensure the manual switch is positioned to the downward setting labeled “Respirable.” If using 

gravimetric filter insert into cassette. Flow can be calibrated by using a calibration jar. Remove 

the GS3 Cyclone Adapter (P/N GSA-204) from the GS3 Cyclone. Attach the GS3 Cyclone to the 

underside of the calibration jar lid. Screw lid tightly. Insert Flow Adapter (p/n FA-104) into the 

IOM Inlet. Attached one tube from the Calibration jar to the Flow Adapter (P/N FA-104). Attach 

the other tube to the flow meter. Select “Run” from the main menu to activate air sampling 

pump. Using the trim stick, adjust the flow on the “Respirable Flow Adjust” located to the left of 

the switch. Adjust to 

2.75 LPM. 

STEP 3

  

Perform Clean Air Zero 

Note: Clean Air Zero needs to be preformed every time the user switches between 

Respirable and Inhalable size fractions.

Insert Flow Adapter (p/n FA-104) into IOM Inlet. Attach tubing from calibration jar to FA-104. 

Attach another tube to a flow meter. 

Remove GSA-204 

GS Cyclone Adapter 

HD-1004 

Sensor

GS Cyclone 

(Purchased separately)

Filter  

Cassette

Flow Adapter 

(P/N FA-104)

Position the switch to the desired 

size fraction. Remove GS Cyclone, 

if attached. Insert the zeroing filter 

(p/n ZA-202A) into the IOM Inlet. 

Select Respirable or Inhalable. Select 

“Run” from the main menu to activate 

air sampling pump. Run monitor 

for 5 minutes to stabilize baseline. 

Select “Enter” to stop sampling. 

Using arrows, select “Auto Zero.” 

When complete, remove the Zeroing 

filter (p/n ZA-202A). To begin an air 

sampling session, select “Run” from 

Main Menu. Select “Overwrite” if you 

do not want to save previous data.

Insert the 

rubber stopper

 of the zeroing 

filter into the 

IOM adapter. 

Connect to 

HD-1004-S Flow Adapter. 

Connect to 

Flow Meter