Step 4:

  

Air Sampling

 

After performing clean air zero switch should already be in the desired position. If sampling for 

Respirable, insert the RS-104 [comprised of the GS3 Cyclone and the GS3 Cyclone Adapter 

(p/n GSA-204)] into IOM inlet at bottom of optical sensor. If sampling for Inhalable, remove 

the RS-104, if attached. Choose the appropriate size fraction on the instrument. Select Enter → 

Special Functions → System Options → Extended Options → Size → Respirable or Inhalable

Environmental Devices Corporation

Fieldstone Industrial Park

4 Wilder Drive Bldg. #15

Plaistow, NH 03865 USA

1(800) 234-2589 or 1(603) 378-2112

Sales

ext.20

[email protected]

Technical Support

ext.14

[email protected]

Visit our website at www.hazdust.com

Insert GS3 Cyclone into IOM inlet at bottom 

of sensor or position cyclone between sensor 

and sensor bracket for a locked fitting. 

Remove Cyclone for Inhalable air sampling.

An “R” will appear when Respirable is 

selected in the menu option. An “I” will 

appear when Inhalable is selected. 

Quick Start Guide

The HD-1004-S is a Real-Time Personal Particulate Air Monitoring Instrument 
with the ability to monitor for Respirable and Inhalable size fractions. When in 
the Inhalable setting, the flow rate should be set to 2.0 LPM and factory dust 
calibration is in accordance with the NIOSH 0500 dust method. When in the 
Respirable setting, the flow rate should be set at 2.75 LPM and used with SKC’s 
Respirable Cyclone and factory dust calibration is in accordance with the NIOSH 
0600 dust Method. When the user has set flow rates the instrument can be used 
to easily switched between Inhalable and Respirable size fractions. The flow rate 
and dust calibration is automatically adjusted. 