2.2

 

Menu and Configuration Settings 

By pressing the temperature button 

 

for 

13 

seconds, the unit enters in to 

Menu status

. There 

are six menu  items by pressing the temperature  button  shortly to 

 

loop  switching  between 

AIR, N2, LX.X, HX.X, A ON/AOFF and E (exit). 
The menu items are described in the following table. 

Menu Items 

Functional Description 

AIR 

“AIR” means Ambient Air Calibration. User presses

 the 

power button 

  

to implement 

Ambient Air 

Calibration 

(300~600ppm). 

N2 

“N2”

 means Nitrogen Calibration. User presses the power 

button 

  

to implement 

Nitrogen 

Calibration 

(0~200ppm). 

H0.5/H0.5T/H1.0/ 
H1.5/H3.0 

Current Defaults are L.0.5 H1.0 

High Alarm Threshold 

setup. 

User presses the power button 

  

to switch the high alarm 

threshold: H0.5 (5000 ppm), H0.5A (5000 ppm), H1.0 (10000 

ppm), H1.5 (15000 

ppm), and H3.0 (30000 ppm). 

L0.5/L0.5A/L1.0/L1.5/L3.0 

Low Alarm Threshold 

setup. User presses the power button   

to switch the low alarm threshold: L0.5 (5000 ppm), L0.5T (5000 
ppm TWA), L1.0 (10000 ppm), L1.5 (15000 ppm), 

and L3.0 (30000 ppm). 

A ON/ AOFF 

Setting 

Man down alarm 

function on/off. User presses the 

power button   

  

to switch “A ON” (allowing

 Man down 

alarm) or “AOFF”

 

(prohibiting Man down alarm). 

User presses the temperature 

button 

 

to exit the menu 

status. 

Содержание ESSAN-10

Страница 1: ...User Manual R1 0 ESSAN 10 Pocket CO2 Meter Distributed Supported By Edaphic Scientific Pty Ltd www edaphic com au info edaphic com au Ph 1300 430 928 ...

Страница 2: ...ts INTRODUCTION 3 FEATURES 3 MONITOR 3 LCD DISPLAY 4 OPERATION 4 MENU CONFIGURATION 5 Calibration 6 Alarm Threshold 6 MAN DOWN FEATURE 7 MAINTENANCE 7 SPECIFICATIONS 8 TROUBLESHOOTING 9 WARRANTY 9 CONTACT US 9 ...

Страница 3: ...udible visual strobe and vibrating alarms Large LCD display Man down alarm leveraging accelerometer technology Rugged design with protective rubber enclosure Heavy duty metal clip Front facing sensor unit Rechargeable Li ion Battery 4 2v 1500mAh 18 hours percharge Black Micro USB cable and Wall USB charger Dual calibration methods Ambient Air 300 600ppm and Nitrogen 0 200ppm Automatic atmospheric ...

Страница 4: ...n the monitor off or change the C F setting 1 Power button 1 1 When the Monitor is turned off press to turn on the unit 1 2 When the Monitor is turned on press for 3 seconds to turn off theunit When the unit is first turned on it performs 5 seconds countdown for Monitor warm up then enters normal display with current CO2 temperature and humidity readings displayed The monitor starts taking measure...

Страница 5: ...sesthe power button to implement Nitrogen Calibration 0 200ppm H0 5 H0 5T H1 0 H1 5 H3 0 Current Defaults are L 0 5 H1 0 High Alarm Threshold setup User presses thepower button to switch the high alarm threshold H0 5 5000 ppm H0 5A 5000 ppm H1 0 10000 ppm H1 5 15000 ppm and H3 0 30000 ppm L0 5 L0 5A L1 0 L1 5 L3 0 Low Alarm Threshold setup User presses the power button to switch the low alarm thre...

Страница 6: ...ure Limit high alarm The alarm thresholds can be changed dependent on the user to L0 5 5000 ppm L0 5T 5000 ppm TWA L1 0 10000 ppm L1 5 15000 ppm or L3 0 30000 ppm If the measured CO2 value exceeds the define Low Alarm threshold the alarm LED will flash red at 2Hz and the audio alarm de de de de will sound If the CO2 level exceeds the High Alarm threshold the LED will flash faster and the audio ala...

Страница 7: ...ettings to change settings 7 Reset Button Users can reset the unit by pushing a reset button through a hole on back of shell MAINTENANCE Calibration The ESSAN 10 comes pre calibrated from the factory However the CO2 sensor should be calibrated at least once a year or as described in your company s safety procedures You can perform the calibration yourself or you can return it to us for factory cal...

Страница 8: ... Pressure Dependence 0 13 of reading per mmHg Response Time About 2 min for 90 of step change Warm up Time 5 seconds at 22 F IAQ LED Display Green 800ppm Yellow 800 1200ppm Red 1200ppm Measurement interval 2 seconds Temperature Sensor Specifications Temperature Range 50 140 F 10 60 C display Display Resolution 0 1 F 0 1 C Display Options F C switchable Accuracy 0 9 F 0 5 C Response Time 5 30 secon...

Страница 9: ... turn on contact support Slow response Power cycle Verify Unit Readings do not change Confirm the monitor has sufficient charge WARRANTY The monitor comes with a one 1 year warranty starting from the date the monitor was shipped to the buyer CONTACT US Distributed Supported By Edaphic Scientific Pty Ltd www edaphic com au info edaphic com au Ph 1300 430 928 ...

Отзывы: