![EDAN SE-101 Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/edan/se-101/se-101_user-manual_2357144082.webp)
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG
- 73 -
6.6.5.1 Editing the HRV Data on the Analysis Screen
Figure 6-21 Analysis Screen of HRV
Designation
Definition
Sampling time
Set sampling time
Total Beat
Total beat number during the measuring course
Heart Rate
Heart rate
Average RR interval
Average RR interval
Max RR interval
Maximum RR interval
Min RR interval
Minimum RR interval
Max/Min
Ratio of Maximum RR interval to Minimum RR interval
SDNN
Standard Deviation of Normal to Normal Intervals
RMSSD
Root Mean Square Successive Difference
NN50
(the total beat number)
The number of duration difference that is more than 50ms
between the adjacent NN durations.
PNN50
(unit:
per centum)
NN50 divide the total NN number